Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Naonori Hu und 13 weiteren Autor:innen. Beteiligte Einrichtung: Kansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Osaka, Japan. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy, Radiation Protection, Neutrons, Lithium. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI.
Abstract (Originalsprache)
Background
Boron neutron capture therapy (BNCT) enables selective tumor irradiation by exploiting the high‐linear energy transfer particles generated from neutron interactions with 10 B atoms. BNCT has been approved as an insurance‐covered medical treatment for recurrent head and neck cancer in Japan. Unlike photon radiotherapy, neutrons that come out of the collimator have an angular distribution. Therefore, it is necessary to keep the distance between the collimator and the patient as short as possible. However, for head and neck cancer treatments, patient anatomy often limits proximity to the collimator, creating an unwanted air gap. This ultimately increases the neutron exposure to surrounding healthy tissue.
Purpose
To develop a freely deformable LiF‐polyethylene neutron shield and assess its impact on neutron/gamma attenuation and clinical organ at‐risk sparing in head and neck BNCT.
Methods
A freely deformable neutron shielding device was constructed using polyethylene beads loaded with lithium fluoride encapsulated in a vacuum‐sealed cushion. Neutron and gamma‐ray attenuation were measured in a water phantom under clinical conditions using an accelerator‐based BNCT system (NeuCure®, Kansai BNCT Medical Center). Measurements were compared with a solid LiF‐polyethylene block and validated through Monte Carlo–based simulations in a commercial treatment planning system. Three representative head and neck cases were further simulated to assess clinical dosimetric effects.
Results
The deformable shielding device reduced the thermal neutron flux by approximately 50%, compared with 60% for the solid LiF‐polyethylene block. Simulated head and neck treatments demonstrated significant OAR dose reductions (up to 46.6% in pharyngeal mucosa D 50% ) without compromising tumor dose coverage ( D 80% ≥ 20 Gy‐eq). Treatment delivery times were minimally affected (< 2 min difference) across all plans. A 5 mm positional perturbation analysis showed ≤ 0.5 Gy‐eq variation in GTV D min and pharyngeal mucosa D 50 and D max .
Conclusions
The freely deformable LiF‐based neutron shielding device effectively attenuated stray neutron dose while maintaining target coverage in BNCT. Its adaptability and reusability make it a practical adjunct for patient‐specific dose optimization in clinical BNCT applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Kansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Osaka, Japan.
- Particle Radiation Oncology Research Center, Kyoto University Institute for Integrated Radiation and Nuclear Science, Osaka, Japan.
- BNCT Joint Clinical Institute, Osaka Medical and Pharmaceutical University, Osaka, Japan.
- Department of Radiology, Kyoto Prefectural University of Medicine, Kyoto, Japan.
- Department of Radiation Oncology, Osaka Medical and Pharmaceutical University Hospital, Osaka, Japan.
- Department of Otorhinolaryngology, Osaka Medical and Pharmaceutical University Hospital, Osaka, Japan.
Förderung
- JSPS KAKENHI
Zitieren
APA
Hu, N., Kakino, R., Sasaki, A., Nojiri, M., Akita, K., Yoshikawa, S., Kohigashi, Y., Yoshino, Y., Takeno, S., Aihara, T., Takata, T., Tanaka, H., Nihei, K., & Ono, K. (2026). Development and dosimetric evaluation of a freely deformable 6 Li‐based neutron shield for boron neutron capture therapy. Medical Physics. https://doi.org/10.1002/mp.70319
Vancouver
Hu N, Kakino R, Sasaki A, Nojiri M, Akita K, Yoshikawa S, et al. Development and dosimetric evaluation of a freely deformable 6 Li‐based neutron shield for boron neutron capture therapy. Medical Physics. 2026. doi:10.1002/mp.70319
BibTeX
@article{Hu2026,
author = {Hu, Naonori and Kakino, Ryo and Sasaki, Akinori and Nojiri, Mai and Akita, Kazuhiko and Yoshikawa, Syuushi and Kohigashi, Yasushi and Yoshino, Yuki and Takeno, Satoshi and Aihara, Teruhito and Takata, Takushi and Tanaka, Hiroki and Nihei, Keiji and Ono, Koji},
title = {Development and dosimetric evaluation of a freely deformable 6 Li‐based neutron shield for boron neutron capture therapy},
journal = {Medical Physics},
year = {2026},
doi = {10.1002/mp.70319}
}
EUVIMED