Look‐up table correction for beam hardening‐induced signal of clinical dark‐field chest radiographs
Medical Physics · 2026-02 · 53 (2)
DOI: 10.1002/mp.70307 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 31.01.2026. Verfasst von Maximilian E Lochschmidt und 6 weiteren Autor:innen. Beteiligte Einrichtung: Chair of Biomedical Physics, Department of Physics, TUM School of Natural Sciences, Technical University of Munich, Garching, Germany. Hauptthemen laut MEDLINE: Radiography, Thoracic, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Research Council (ERC Synergy Grant SmartX), Center for Advanced Laser Applications (CALA), Technical University of Munich - Institute for Advanced Study.
Abstract (Originalsprache)
Background
The microstructure of material at a length scale leads to ultra‐small‐angle scattering of X‐rays, which typically occurs, e.g., for lung tissue or some plastic foams. When using an interferometer, this effect alters the visibility of the fringe pattern, which can be detected and resolved by the detector. Thus, the ultra‐small‐angle scattering can be represented as a dark‐field image. For a polychromatic source, the hardening of the source spectrum changes visibility as well, generating an additional fake dark‐field signal by the attenuation of the material on top of the real ultra‐small‐angle scatter‐related dark‐field signal. Consequently, even homogeneous materials without microstructure typically exhibit a change in visibility.
Purpose
The objective of this study is to develop a fast, simple, and robust method to correct dark‐field signals and bony structures present due to beam hardening on dark‐field chest radiographs of study participants.
Methods
The method is based on calibration measurements and image processing. BH by bones and soft tissue is modeled by aluminum and water, respectively, which have no microstructure and thus only generate an artificial dark‐field signal. Look‐up tables were then created for both. By using a weighted mean of these, forming a single LUT, and using the attenuation images, the artificial dark‐field signal and thus the bone structures present are reduced for study participants.
Results
It was found that applying a correction using a weighted LUT leads to a significant reduction of bone structures in the dark‐field image. The weighting of the aluminum component has a substantial impact on the degree to which bone structures remain visible in the dark‐field image. Furthermore, a large negative bias in the dark‐field image–dependent on the aluminum weighting–was successfully corrected.
Conclusions
BH‐induced signal in the dark‐field images was successfully reduced using the method described. The choice of aluminum weighting to suppress rib structures, as well as the selection of bias correction, should be evaluated based on the specific clinical question.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Chair of Biomedical Physics, Department of Physics, TUM School of Natural Sciences, Technical University of Munich, Garching, Germany.
- Munich Institute of Biomedical Engineering, Technical University of Munich, Garching, Germany.
- Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM Klinikum, Technical University of Munich (TUM), München, Germany.
- TUM Institute for Advanced Study, Technical University of Munich, Garching, Germany.
- Philips Innovative Technologies, Hamburg, Germany.
ORCID-Profile: Maximilian E Lochschmidt, Theresa Urban, Lennard Kaster, Rafael C Schick, Thomas Koehler, Daniela Pfeiffer, Franz Pfeiffer
Förderung
- European Research Council (ERC Synergy Grant SmartX)
- Center for Advanced Laser Applications (CALA)
- Technical University of Munich - Institute for Advanced Study
- The Free State of Bavaria under the Excellence Strategy of the Federal Government and the Länder
Zitieren
APA
Lochschmidt, M. E., Urban, T., Kaster, L., Schick, R. C., Koehler, T., Pfeiffer, D., & Pfeiffer, F. (2026). Look‐up table correction for beam hardening‐induced signal of clinical dark‐field chest radiographs. Medical Physics, 53(2). https://doi.org/10.1002/mp.70307
Vancouver
Lochschmidt ME, Urban T, Kaster L, Schick RC, Koehler T, Pfeiffer D, et al. Look‐up table correction for beam hardening‐induced signal of clinical dark‐field chest radiographs. Medical Physics. 2026;53(2). doi:10.1002/mp.70307
BibTeX
@article{Lochschmidt2026,
author = {Lochschmidt, Maximilian E and Urban, Theresa and Kaster, Lennard and Schick, Rafael C and Koehler, Thomas and Pfeiffer, Daniela and Pfeiffer, Franz},
title = {Look‐up table correction for beam hardening‐induced signal of clinical dark‐field chest radiographs},
journal = {Medical Physics},
year = {2026},
volume = {53},
number = {2},
doi = {10.1002/mp.70307}
}
EUVIMED