Exposure control at the first dark-field chest radiography demonstrator system

M E Lochschmidt, M Frank, K Willer, H Bast, L Kaster, T Urban, W Noichl, R Schick, F De Marco, T Koehler, I Maack, G S Zimmermann und 3 weitere

Physica Medica · 2025

DOI: 10.1016/j.ejmp.2025.105013 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physica Medica, 2025. Verfasst von M E Lochschmidt und 14 weiteren Autor:innen. Beteiligte Einrichtung: Chair of Biomedical Physics, Department of Physics, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany; Munich I…. Hauptthemen laut MEDLINE: Radiography, Thoracic. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Institute for Advanced Study, European Research Council, Deutsche Forschungsgemeinschaft.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: A first demonstrator system for clinical dark-field chest radiography was recently constructed. Due to its scanning acquisition, conventional exposure regulation systems are not possible. This study presents the adaptation of exposure to each patient individually, using a conventional radiograph as a reference. Additionally, an alternative approach for individual exposure planning based on the patient's body mass index (BMI) is proposed. METHODS: Using exposure settings from a conventional radiography system with automatic exposure control (AEC), a patient-specific equivalent attenuator thickness was calculated with polyoxymethylene (POM) as a surrogate material. The tube current at the dark-field system was adapted to reach the target detector dose based on this thickness. This approach was verified with both phantom measurements and patient examinations. The correlation between patient body parameters (weight, BMI, upper bust girth, and under-bust girth) and desired tube current was also evaluated. RESULTS: A calibration curve was found to transfer individual exposure settings from a conventional device to the dark-field system. Achieved detector doses for both phantom and patient examinations were within permitted ranges. A strong correlation was found between each body parameter and the desired tube current, with BMI showing the strongest correlation (r=0.87,p

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiography, ThoracicHumansPhantoms, ImagingRadiation DosageCalibrationBody Mass Index

Stichwörter der Autor:innen

Clinical dark-field chest radiographyExposure adaptionExposure controlExposure planningGrating based imagingX-ray radiography

Autor:innen und Institutionen

  • Chair of Biomedical Physics, Department of Physics, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany; Munich Institute of Biomedical Engineering, Technical University of Munich, 85…
  • Philips Innovative Technologies, 22335, Hamburg, Germany; TUM Institute for Advanced Study, Technical University of Munich, 85748 Garching, Germany.
  • Philips Medical Systems DMC, 22335, Hamburg, Germany.
  • Division of Respiratory Medicine, Department of Internal Medicine I, School of Medicine and Health, TUM Klinikum, Technical University of Munich (TUM), 81675 München, Germany.
  • Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM Klinikum, Technical University of Munich (TUM), 81675 München, Germany; TUM Institute for Advanced Study, Technical University of…
  • Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM Klinikum, Technical University of Munich (TUM), 81675 München, Germany.

Förderung

  • Institute for Advanced Study
  • European Research Council
  • Deutsche Forschungsgemeinschaft
  • Bundesministerium für Bildung und Forschung
  • Karlsruhe Institute of Technology
  • Technische Universität München

Zitieren

APA

Lochschmidt, M. E., Frank, M., Willer, K., Bast, H., Kaster, L., Urban, T., Noichl, W., Schick, R., De Marco, F., Koehler, T., Maack, I., Zimmermann, G. S., Pfeiffer, D., Renger, B., & Pfeiffer, F. (2025). Exposure control at the first dark-field chest radiography demonstrator system. Physica Medica. https://doi.org/10.1016/j.ejmp.2025.105013

Vancouver

Lochschmidt ME, Frank M, Willer K, Bast H, Kaster L, Urban T, et al. Exposure control at the first dark-field chest radiography demonstrator system. Physica Medica. 2025. doi:10.1016/j.ejmp.2025.105013

BibTeX
@article{Lochschmidt2025,
  author = {Lochschmidt, M E and Frank, M and Willer, K and Bast, H and Kaster, L and Urban, T and Noichl, W and Schick, R and De Marco, F and Koehler, T and Maack, I and Zimmermann, G S and Pfeiffer, D and Renger, B and Pfeiffer, F},
  title = {Exposure control at the first dark-field chest radiography demonstrator system},
  journal = {Physica Medica},
  year = {2025},
  doi = {10.1016/j.ejmp.2025.105013}
}
Kostenlos starten →