EUVIMEDΑνακαλύψτε το EUVIMED
Open access · CC BY 4.0

Image reconstruction and elongation artifact reduction for a dual‐panel dedicated prostate PET scanner

Abdollah Saberi Manesh iD, Mehdi Amini iD, Yazdan Salimi iD, Katayoun Doroud, Crispin Williams, Themistoklis Williams, Hossein Arabi iD, Habib Zaidi iD

Medical Physics · 2026

DOI: 10.1002/mp.70298 ↗

At a glance

Journal article in Medical Physics, 2026. Written by Abdollah Saberi Manesh and 7 co-authors. Institution: Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, Geneva, Switzerland. Major MEDLINE topics: Positron-Emission Tomography, Image Processing, Computer-Assisted, Prostate, Artifacts. 40 references. The full text is freely available (open access, CC BY 4.0). Funded by: Eurostars program of the European Commission, Private Foundation of Geneva University Hospitals.

ΠερίληψηSubjectsΒιβλιογραφικά στοιχείαCiteRelated studies

Abstract (original language)

Background

The development of PET scanners dedicated to high temporal and spatial resolution organ‐specific imaging is an active research area, motivated by the need for cost reduction, improved lesion detectability and quantification in specific clinical scenarios, as well as by ongoing hardware and software innovations.

Purpose

This study investigates and compares various image reconstruction strategies for a dual‐panel prostate‐dedicated PET scanner (ProVision), which features four‐layered dual‐readout time‐of‐flight depth‐of‐interaction detectors and a 22‐position acquisition protocol to improve angular coverage. Materials and methods A list‐mode MLEM algorithm with multi‐ray modeling was developed and optimized using a scaled NEMA image quality phantom to determine optimal number of rays and iterations. These parameters were then used to reconstruct data from both simulation and experimental acquisitions, including an anthropomorphic pelvis phantom, named Adam‐PETer. Four reconstruction approaches were evaluated: classical MLEM; MLEM with embedded shift‐variant point spread function (PSF) modeling; a hybrid list‐mode reconstruction; and a Swin‐UNETR‐based deep learning model applied as a post‐reconstruction enhancement to MLEM images. Performance was assessed using contrast recovery coefficient (CRC), contrast‐to‐noise ratio (CNR), and contrast‐to‐noise consistency (CNC), on both a scaled NEMA phantom and an experimental anthropomorphic phantom.

Results

In the scaled NEMA phantom simulation, the Swin‐based method yielded the highest CNR and CNC, especially for the smallest spheres, thereby outperforming both standard MLEM and the hybrid algorithm. In the Adam‐PETer experimental prostate phantom, the CNR was 10.43 for MLEM, 14.48 for Hybrid, and 13.85 for Swin for the larger lesion (10 mm). The CNR values were 2.28, 3.03, and 4.35, respectively, for the smaller lesion (8 mm). CNC values also varied across methods, with Swin achieving the best result for the smaller lesion. These findings indicate that model‐based and learned methods offer complementary strengths depending on lesion contrast and size.

Conclusion

The PET scanner‐adapted reconstruction combined with deep learning refinement improves image quality in dedicated, limited‐angle PET systems.

Subjects

MeSH terms (MEDLINE)

Positron-Emission TomographyImage Processing, Computer-AssistedProstateArtifactsPhantoms, ImagingMaleHumansAlgorithms

Author keywords

artifactdeep‐learningdual‐panelorgan‐specific PETprostatereconstruction

Authors and institutions

  • Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, Geneva, Switzerland.
  • Picotech, St Genis Pouilly, France.
  • Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center, Groningen, The Netherlands.
  • Department of Nuclear Medicine, University of Southern Denmark, Odense, Denmark.
  • University Research and Innovation Center, Óbuda University, Budapest, Hungary.

ORCID profiles: Abdollah Saberi Manesh, Mehdi Amini, Yazdan Salimi, Hossein Arabi, Habib Zaidi

Funding

  • Eurostars program of the European Commission
  • Private Foundation of Geneva University Hospitals

Cite

APA

Saberi Manesh, A., Amini, M., Salimi, Y., Doroud, K., Williams, C., Williams, T., Arabi, H., & Zaidi, H. (2026). Image reconstruction and elongation artifact reduction for a dual‐panel dedicated prostate PET scanner. Medical Physics. https://doi.org/10.1002/mp.70298

Vancouver

Saberi Manesh A, Amini M, Salimi Y, Doroud K, Williams C, Williams T, et al. Image reconstruction and elongation artifact reduction for a dual‐panel dedicated prostate PET scanner. Medical Physics. 2026. doi:10.1002/mp.70298

BibTeX
@article{SaberiManesh2026,
  author = {Saberi Manesh, Abdollah and Amini, Mehdi and Salimi, Yazdan and Doroud, Katayoun and Williams, Crispin and Williams, Themistoklis and Arabi, Hossein and Zaidi, Habib},
  title = {Image reconstruction and elongation artifact reduction for a dual‐panel dedicated prostate PET scanner},
  journal = {Medical Physics},
  year = {2026},
  doi = {10.1002/mp.70298}
}
Start for free →