Minibeam‐pLATTICE: A novel proton LATTICE modality using minibeams
Medical Physics · 2026-01 · 53 (1)
DOI: 10.1002/mp.70250 ↗At a glance
Journal article in Medical Physics, 2026-01. Written by Nimita Shinde and 3 co-authors. Institution: Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA. Major MEDLINE topics: Proton Therapy. Cited 8 times so far (Crossref); 49 references. Funded by: NIH HHS, National Institutes of Health.
Abstract (original language)
Background
LATTICE, a form of spatially fractionated radiation therapy (SFRT) that delivers high‐dose peaks and low‐dose valleys within the target volume, has been clinically utilized for treating bulky tumors. However, its application to small‐to‐medium‐sized target volumes remains challenging due to beam size limitations.
Purpose
To address the challenge of applying LATTICE radiation therapy to small‐to‐medium‐sized targets, this work proposes a novel proton LATTICE (pLATTICE) modality using minibeams, namely minibeam‐pLATTICE, that can extend the LATTICE approach for small‐to‐medium target volumes.
Methods
Three minibeam‐pLATTICE methods are introduced. (1) M0: a fixed minibeam aperture orientation (e.g., 0°) for all beam angles; (2) M1: alternated minibeam aperture orientations (e.g., between 0° and 90°), for consecutive beam angles; (3) M2: multiple minibeam aperture orientations (e.g., 0° and 90°) for each beam angle. The purpose of M1 or M2 is to correct anisotropic dose distribution at lattice peaks due to the planar spatial modulation of minibeams. For each minibeam‐pLATTICE method, an optimization problem is formulated to optimize dose uniformity in target peaks and valleys, as well as dose‐volume‐histogram‐based objectives. This optimization problem is solved using iterative convex relaxation and alternating direction method of multipliers (ADMM).
Results
Three minibeam‐pLATTICE methods are validated to demonstrate the feasibility of minibeam‐pLATTICE for two clinical head‐and‐neck (HN), one abdominal, and one brain case. The advantages of this modality over conventional beam (CONV) pLATTICE are evaluated by comparing peak‐to‐valley dose ratio (PVDR) and dose delivered to organs at risk (OAR). All three minibeam‐pLATTICE modalities achieved improved plan quality compared to CONV, with M2 yielding the best results. For instance, in one HN case, the following improvements were observed: PVDR increased to 3.73 (M2), compared to 3.27 (CONV), 3.72 (M0), and 3.49 (M1), while the mean dose to the mandible was reduced to 0.18 Gy (M2), compared to 0.33 Gy (CONV), 0.17 Gy (M0), and 0.14 Gy (M1).
Conclusions
A novel minibeam‐pLATTICE modality is proposed that can generate lattice dose patterns for small‐to‐medium target volumes, which are not achievable with conventional pLATTICE due to beam size limitations. Peak dose anisotropy due to 1D planar minibeam apertures is corrected through inverse treatment planning with alternating or multiple minibeam apertures per beam angle.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
- Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas, USA.
Funding
- NIH HHS
- National Institutes of Health
Cite
APA
Shinde, N., Zhang, W., Lin, Y., & Gao, H. (2026). Minibeam‐pLATTICE: A novel proton LATTICE modality using minibeams. Medical Physics. https://doi.org/10.1002/mp.70250
Vancouver
Shinde N, Zhang W, Lin Y, Gao H. Minibeam‐pLATTICE: A novel proton LATTICE modality using minibeams. Medical Physics. 2026. doi:10.1002/mp.70250
BibTeX
@article{Shinde2026,
author = {Shinde, Nimita and Zhang, Weijie and Lin, Yuting and Gao, Hao},
title = {Minibeam‐pLATTICE: A novel proton LATTICE modality using minibeams},
journal = {Medical Physics},
year = {2026},
doi = {10.1002/mp.70250}
}
EUVIMED