Development of a customizable model for spectral photon‐counting detector CT
Medical Physics · 2025
DOI: 10.1002/mp.70195 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Mridul Bhattarai und 4 weiteren Autor:innen. Beteiligte Einrichtung: Center for Virtual Imaging Trials, Department of Radiology, Duke University, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Photons, Tomography, X-Ray Computed, Models, Theoretical. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIH HHS, NHLBI NIH HHS, NIBIB NIH HHS.
Abstract (Originalsprache)
Background
Photon‐counting detector CT (PCD‐CT) is a new CT technology that offers enhanced spatial and spectral imaging performances. As a new technology, conditioning and qualifying its precise performance can benefit from a comprehensive framework to evaluate task‐generic and task‐specific image qualities.
Purpose
To develop and validate a customizable and physics‐informed simulation framework capable of modeling spatio‐energetic detector responses for various PCD designs, integrate it into a virtual imaging framework, and demonstrate its applicability in clinically relevant imaging tasks.
Methods
A customizable simulation model, DukeCounter, was developed to replicate real PCD‐CT systems. Photon transport and crosstalk in PCDs were modeled using Monte Carlo simulations, and charge sharing was implemented using an analytical Gaussian charge cloud model. The fundamental interactions in PCDs, including photoelectric absorption, Compton and fluorescence x‐ray scatterings, charge cloud formation, and charge diffusion and repulsion, were modeled. Spatio‐energetic detector responses were generated for face‐on CdTe‐, CZT‐, GaAs‐, and edge‐on Si‐based PCDs. These responses, combined with standardized scanner parameters, were integrated into a CT simulator to create virtual DukeCounter PCD‐CT scanners. The framework was benchmarked against experimental data from a clinical CdTe‐based PCD‐CT scanner across three dose levels. To demonstrate its utility, three pilot studies were conducted using a computational ACR phantom for task‐generic image quality assessment, an XCAT model with bronchitis and emphysema for COPD biomarker extraction, and an XCAT with liver lesions for lesion detectability analysis.
Results
The simulated charge cloud size increased with energy and was more pronounced in Si due to its low atomic number. The detector response across a 3 × 3‐pixel neighborhood varied with PCD material, design, and energy threshold settings. Validation results demonstrated strong agreement between simulated and real ACR images. For the 20‐keV‐threshold images, the mean relative difference (MRD) in f 50 of MTF was 4.15% ± 1.21 for air and 2.54% ± 2.08 for bone, and the MRD in f av of NPS was 0.83% ± 0.97. The MRDs in noise magnitude were 2.65% ± 1.68, 3.05% ± 1.97, and 2.78% ± 1.79 for the 20‐keV‐threshold, 65‐keV‐threshold, and 70‐keV‐VMI images, respectively. The MRDs in CT number for the same image types were 0.03% ± 0.03, 0.11% ± 0.09, 0.11% ± 0.05 for air, and 1.85% ± 0.20, 1.84% ± 0.55, 0.50% ± 0.36 for polyethylene. DukeCounter‐generated images showed that task‐generic and task‐specific image qualities were influenced by PCD materials, designs, and energy threshold settings. GaAs‐based DukeCounter exhibited the highest image noise, the largest error in COPD biomarker quantification, and the lowest performance in liver lesion detection, under consistent acquisition and reconstruction settings.
Conclusions
A customizable, modular simulation framework was developed to model spatio‐energetic detector responses for various PCD materials and designs. The detector responses were integrated into a CT simulation pipeline to build DukeCounter PCD‐CT systems. The framework's utility was demonstrated through task‐specific assessments of image quality and clinical performance of DukeCounter systems using XCAT phantoms. This approach enables systematic PCD‐CT design evaluation and optimization, supporting translational research in medical imaging by reducing the cost, time, and radiation burden of physical experiments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Virtual Imaging Trials, Department of Radiology, Duke University, Durham, North Carolina, USA.
- Center for Virtual Imaging Trials (CVIT), Duke University, Durham, North Carolina, USA.
- Department of Radiology, Duke University School of Medicine, Durham, North Carolina, USA.
Förderung
- NIH HHS
- NHLBI NIH HHS
- NIBIB NIH HHS
- National Science Foundation
Zitieren
APA
Bhattarai, M., Panta, R. K., Segars, W. P., Abadi, E., & Samei, E. (2025). Development of a customizable model for spectral photon‐counting detector CT. Medical Physics. https://doi.org/10.1002/mp.70195
Vancouver
Bhattarai M, Panta RK, Segars WP, Abadi E, Samei E. Development of a customizable model for spectral photon‐counting detector CT. Medical Physics. 2025. doi:10.1002/mp.70195
BibTeX
@article{Bhattarai2025,
author = {Bhattarai, Mridul and Panta, Raj Kumar and Segars, W Paul and Abadi, Ehsan and Samei, Ehsan},
title = {Development of a customizable model for spectral photon‐counting detector CT},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.70195}
}
EUVIMED