Joint range‐modulator and spot optimization for Bragg‐peak proton FLASH radiotherapy
Medical Physics · 2025-12 · 52 (12)
DOI: 10.1002/mp.70171 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 28.11.2025. Verfasst von Jiayue Han und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas, USA. Hauptthemen laut MEDLINE: Proton Therapy, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Background
Ultra‐high‐dose‐rate (UHDR) radiation therapy has demonstrated promising potential in reducing toxicity to organs‐at‐risk (OARs). Proton therapy is uniquely positioned to deliver UHDR by leveraging the Bragg peak in conjunction with patient‐specific range modulators (PSRMs) to generate a spread‐out Bragg peak (SOBP). Existing proton FLASH (pFLASH) treatment planning workflows typically follow a two‐step process: 1) generating a multi‐energy intensity‐modulated proton therapy (IMPT) plan to determine spot weights and 2) subsequently converting this plan into a single‐energy pFLASH delivery using PSRM optimization. However, the intrinsic coupling between spot weight distribution and PSRM design has not been fully investigated, which may limit the achievable dosimetric and radiobiological advantages of pFLASH therapy.
Purpose
This work proposes a novel alternating optimization framework—Joint Range‐Modulator and Spot Optimization (JRSO)—that simultaneously optimizes the PSRM and spot weights to improve the plan quality of conformal pFLASH therapy.
Methods
Compared to the conventional method, JRSO does not require a one‐to‐one correspondence between beam spots and PSRM pins. Proton beam diffusion from the delivery system to the PSRM is modeled by Gaussian function. To achieve better plan quality, starting from an initial solution derived from a conventional IMPT plan, JRSO iteratively updates the PSRM design and spot weights in an alternating manner. This process progressively refines both parameters while ensuring compliance with practical delivery constraints, such as the minimum monitor‐unit (MMU) requirement.
Results
JRSO obtained improved plan quality compared to the conventional method. For example, in a head‐and‐neck (HN) case, JRSO reduced the objective function value from 0.46 to 0.26, lowered the maximum target dose from 117.6% to 107.1%, improved the conformity index from 0.74 to 0.87, and decreased the region‐of‐interest (ROI) effective dose from 6.50 to 6.10 Gy.
Conclusion
A new optimization method JRSO is proposed for conformal pFLASH radiotherapy. It outperforms the conventional approach and may extend the applicability of PSRM to more complex clinical scenarios, particularly those involving misalignments between beam spots and pins. Numerical results demonstrate the robustness and efficiency of the new method.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas, USA.
- Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, China.
- School of Mathematics, Harbin Institute of Technology, Harbin, China.
- Department of Radiation Oncology, University of Texas Southwestern Medical Center, Texas, USA.
Förderung
- NIH HHS
- National Institutes of Health
Zitieren
APA
Han, J., Wang, A., Zhu, Y. N., Li, W., Lin, Y., & Gao, H. (2025). Joint range‐modulator and spot optimization for Bragg‐peak proton FLASH radiotherapy. Medical Physics, 52(12). https://doi.org/10.1002/mp.70171
Vancouver
Han J, Wang A, Zhu YN, Li W, Lin Y, Gao H. Joint range‐modulator and spot optimization for Bragg‐peak proton FLASH radiotherapy. Medical Physics. 2025;52(12). doi:10.1002/mp.70171
BibTeX
@article{Han2025,
author = {Han, Jiayue and Wang, Aoxiang and Zhu, Ya-Nan and Li, Wangyao and Lin, Yuting and Gao, Hao},
title = {Joint range‐modulator and spot optimization for Bragg‐peak proton FLASH radiotherapy},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {12},
doi = {10.1002/mp.70171}
}