Development of an adjustable surface coil for low‐field magnetic resonance imaging‐guided breast interventions

Qingyun Liu, Xueyan Song, Qing Zhang, Kecheng Yuan, Lin Chen, Bensheng Qiu

Medical Physics · 2025

DOI: 10.1002/mp.70153 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Qingyun Liu and 5 co-authors. Institution: Department of Electronic Engineering and Information Science, Medical Imaging Center, University of Science and Technology of China, Hefei, Anhui, Ch…. Major MEDLINE topics: Magnetic Resonance Imaging, Breast, Breast Neoplasms, Surgery, Computer-Assisted. Cited 1 times so far (Crossref); 41 references. Funded by: National Key Research and Development Program of China.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

As breast cancer becomes the most common malignant tumor in women, magnetic resonance imaging (MRI)‐guided breast interventions have gradually demonstrated their practical value in clinical practice.

Purpose

Low‐field MRI systems have emerged as a new option for breast intervention platforms due to their high openness, low cost, mobility, and the ability to facilitate real‐time surgical procedures. Here, we designed an adjustable splint‐integrated surface coil combining butterfly and ring units for 0.35 T MRI‐guided breast interventions.

Methods

Electromagnetic simulations were first conducted to analyze the radiofrequency field distribution in the imaging region, with near‐field simulations evaluating the adjustable coil under various configurations. Subsequently, bench tests were performed to analyze the S‐parameter of the coil circuit while measuring both the Q‐factor and Q‐ratio, ensuring compliance with standard electrical performance requirements. Then, phantom imaging quantified MR image signal‐to‐noise ratio (SNR) to compare quality across different configurations and determine the two‐channel adjustable coil design for low‐field MRI‐guided breast interventions. MRI compared the adjustable surface and volume coils using a silicone breast phantom and two volunteers with splint spacing adjusted to 80 or 40 mm based on breast size, while SNR maps visually compared image quality across different configurations. The final ex vivo porcine liver puncture experiment validated the clinical applicability of the two‐channel adjustable surface coil.

Results

The adjustable coil demonstrated high SNR (> 130) in clinically relevant breast compression scenarios (< 40 mm spacing), with adequate performance in the planar dual‐channel configuration (SNR = 108.98) compared to curved designs (SNR = 99.87). And the four‐channel coil overlap decoupling compromised grid accessibility, justifying the final selection of a simplified two‐channel planar design that balances imaging quality and interventional practicality for MRI‐guided breast procedures. The two‐channel adjustable surface coil observably outperformed conventional volume coils, providing SNR improvements ranging from 14% to 43% for both phantoms and breast tissue in MRI‐guided interventions. Additionally, the ex vivo porcine liver puncture experiment validated the clinical feasibility of the complete MRI‐guided interventional workflow utilizing the adjustable coil system, demonstrating 1 mm targeting precision.

Conclusion

This adjustable surface coil is designed to work with the open low‐field MRI system, providing high imaging SNR for real‐time guidance during breast interventional surgeries and holding remarkable potential for extension to diverse MR imaging applications.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance ImagingBreastBreast NeoplasmsSurgery, Computer-AssistedEquipment DesignAnimalsSwineHumansPhantoms, ImagingSignal-To-Noise RatioFemale

Author keywords

SNR mapadjustable coilbreast MRIinterventional splintopen system

Authors and institutions

  • Department of Electronic Engineering and Information Science, Medical Imaging Center, University of Science and Technology of China, Hefei, Anhui, China.
  • Anhui Fuqing Medical Equipment Co., Ltd, Hefei, Anhui, China.

Funding

  • National Key Research and Development Program of China

Cite

APA

Liu, Q., Song, X., Zhang, Q., Yuan, K., Chen, L., & Qiu, B. (2025). Development of an adjustable surface coil for low‐field magnetic resonance imaging‐guided breast interventions. Medical Physics. https://doi.org/10.1002/mp.70153

Vancouver

Liu Q, Song X, Zhang Q, Yuan K, Chen L, Qiu B. Development of an adjustable surface coil for low‐field magnetic resonance imaging‐guided breast interventions. Medical Physics. 2025. doi:10.1002/mp.70153

BibTeX
@article{Liu2025,
  author = {Liu, Qingyun and Song, Xueyan and Zhang, Qing and Yuan, Kecheng and Chen, Lin and Qiu, Bensheng},
  title = {Development of an adjustable surface coil for low‐field magnetic resonance imaging‐guided breast interventions},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.70153}
}
Start for free →