K‐Edge imaging using a clinical dual‐source photon‐counting CT system
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70726 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Martin V Rybertt und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Photons, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers, NIH HHS, NIBIB NIH HHS.
Abstract (Originalsprache)
Background
Photon‐counting CT (PCCT) enables K‐edge imaging, an advanced multi‐material decomposition technique that requires multiple energy bins to perform decomposition unconstrained by physical assumptions. The clinical implementation of K‐edge imaging applications like dual‐contrast imaging could lead to reductions in radiation dose and motion artifacts between sequential scans.
Purpose
To evaluate the feasibility and performance of K‐edge imaging of iodine (I) and gadolinium (Gd) on a clinically available photon‐counting computed tomography (PCCT) system.
Methods
A dual‐source clinical PCCT scanner with four energy thresholds (20, 55, 72, 90 keV) was used to scan phantoms containing single and dual contrast solutions of I and Gd across multiple concentrations (1–10 mg/mL) and radiation doses (1–8 mGy). Multi‐material decomposition was performed using a calibration‐based, image‐domain algorithm to generate material‐specific maps. Quantitative accuracy was assessed using Bland‐Altman analysis and contrast‐to‐noise ratio (CNR), while noise and bias trends were statistically analyzed using non‐parametric tests.
Results
K‐edge imaging was successfully achieved on a clinical PCCT system with accurate decomposition of I and Gd across varying concentrations, solution compositions (single/dual contrast), and dose levels. Quantitative bias was significantly influenced by radiation dose, concentration, and solution composition ( p < 0.0004). Increased radiation dose and contrast concentration improved quantification accuracy, with maximum bias reductions of 0.9 (I) and 0.3 mg/mL (Gd). CNR correlated linearly with concentration ( R 2 > 0.99) and moderately with dose ( R 2 = 0.85–0.94), achieving peak values of 13 (I) and 16 (Gd) at 8 mGy. Dual contrast solutions showed reduced performance compared to single contrast solutions, that is CNR of 5 mg/mL Gd solutions increased by 0.6 per mGy in single contrast solutions while by 0.5 per mGy in dual contrast mixtures. Noise was dependent on dose but not on concentration or solution composition.
Conclusion
This study establishes the feasibility of K‐edge imaging using a clinical PCCT system and demonstrates accurate, simultaneous decomposition of I and Gd across composition of solution. These findings support the clinical translation of K‐edge imaging and highlight PCCT's potential for advanced dual‐contrast and molecular imaging applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Siemens Healthineers, Malvern, Pennsylvania, USA.
- Siemens Healthineers AG, Forchheim, Germany.
- Cleveland Clinic Florida, Weston, Florida, USA.
Förderung
- Siemens Healthineers
- NIH HHS
- NIBIB NIH HHS
- National Institutes of Health
Zitieren
APA
Rybertt, M. V., Liu, L. P., Sahbaee, P., Nowak, T., Mathew, M., Litt, H. I., Cormode, D. P., Dhanaliwala, A. H., & Noël, P. B. (2026). K‐Edge imaging using a clinical dual‐source photon‐counting CT system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70726
Vancouver
Rybertt MV, Liu LP, Sahbaee P, Nowak T, Mathew M, Litt HI, et al. K‐Edge imaging using a clinical dual‐source photon‐counting CT system. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70726
BibTeX
@article{Rybertt2026,
author = {Rybertt, Martin V and Liu, Leening P and Sahbaee, Pooyan and Nowak, Tristan and Mathew, Manoj and Litt, Harold I and Cormode, David P and Dhanaliwala, Ali H and Noël, Peter B},
title = {K‐Edge imaging using a clinical dual‐source photon‐counting CT system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70726}
}
EUVIMED