A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions

Ruizhe Hou, Quanshu Han, Jincheng Zou, Shiqing Zhao, Aili Zhang

Medical Physics · 2025

DOI: 10.1002/mp.18112 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Ruizhe Hou and 4 co-authors. Institution: School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, P.R. China. Major MEDLINE topics: Phantoms, Imaging, Temperature, Catheter Ablation, Radiofrequency Ablation. 46 references. Funded by: National Natural Science Foundation of China, the Ministry of Science and Technology of China, the Interdisciplinary Program of Shanghai Jiao Tong University.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

In catheter‐based radiofrequency ablation (RFA), energy is delivered to heterogeneous thin‐walled tissues to induce therapeutic heating. Variations in electrical and mechanical properties of tissue contents have a great effect on outcomes.

Purpose

The objective of this study is to develop models that replicate tissue heterogeneity and visualize ablation zones for effective evaluation and optimization.

Methods

A tunable thermochromic phantom formulation was developed, consisting of deionized water, acrylamide aqueous solution (Acr/Bis), thermochromic microcapsule powder, Tris‐HCl buffer, ammonium persulfate solution, and N , N , N ', N '‐tetramethylethylenediamine. The color‐change temperature threshold, electrical conductivity, and elastic modulus were characterized with different formulation compositions. A dual‐parameter inverse design method based on concentration‐property regression models was proposed to match the electrical and mechanical properties of tissues simultaneously, which was validated through RFA experiments on phantoms.

Results

The phantoms exhibited an immediate and irreversible color change from indigo to deep pink at 60°C, clearly delineating ablation margins. Electrical conductivity ranged from 0.1892 to 0.5958 S/m and elastic modulus from 6.70 to 138 kPa, effectively encompassing the characteristic properties of diverse lesion tissues. RFA experiments on phantoms mimicking porcine myocardium and chicken breast demonstrated strong agreement with ex vivo results. Ablation of heterogeneous phantoms mimicking atherosclerotic structures revealed significant differences compared to homogeneous models.

Conclusions

A tunable thermochromic phantom formulation was developed, exhibiting a precise color change at 60°C while replicating the electrical and mechanical properties of heterogeneous tissues, enabling direct visualization of ablation boundaries. The phantom offers a robust platform for evaluating catheter‐based RFA in clinically relevant complex tissues, supporting the development of personalized and optimized treatment strategies.

Subjects

MeSH terms (MEDLINE)

Phantoms, ImagingTemperatureCatheter AblationRadiofrequency AblationAnimalsSwineElectric ConductivityChickensCatheters

Author keywords

mechanical propertyradiofrequency ablationthermal therapythermochromictissue‐mimicking phantom

Authors and institutions

  • School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, P.R. China.

Funding

  • National Natural Science Foundation of China
  • the Ministry of Science and Technology of China
  • the Interdisciplinary Program of Shanghai Jiao Tong University

Cite

APA

Hou, R., Han, Q., Zou, J., Zhao, S., & Zhang, A. (2025). A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions. Medical Physics. https://doi.org/10.1002/mp.18112

Vancouver

Hou R, Han Q, Zou J, Zhao S, Zhang A. A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions. Medical Physics. 2025. doi:10.1002/mp.18112

BibTeX
@article{Hou2025,
  author = {Hou, Ruizhe and Han, Quanshu and Zou, Jincheng and Zhao, Shiqing and Zhang, Aili},
  title = {A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.18112}
}
Start for free →