X‐ray multimeter performance under calibration laboratory conditions
Medical Physics · 2025-09 · 52 (9)
DOI: 10.1002/mp.18106 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 03.09.2025. Verfasst von John T Stasko und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Laboratories, Radiation Dosimeters. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Estimating dose delivered to patients for a typical mammographic or radiologic examination requires accurate knowledge of several beam quantities. X‐ray multimeters (XMMs) are compact, solid‐state semiconductor dosimeters that have become common for conventional QA measurements due to their ease of use.
Purpose
In this investigation, the performance of two XMMs in low‐energy x‐ray calibration beams was assessed, and the stability of the manufacturer's calibration over time was evaluated.
Methods
An RTI Piranha and Radcal Accu‐Gold+ with AGMS‐DM+ sensor were used to measure air‐kerma rates and half‐value layers of UW‐MO and UW‐M series calibration beams. Measurement results were compared to reference values collected with standard instruments. The same measurements were repeated every 3–5 months for 2.5 years to evaluate whether the XMMs’ energy response changes over time.
Results
All air‐kerma rate measurements and nearly all half‐value layer measurements were within tolerance of the reference measurements. Both XMMs were satisfactorily stable over the course of the study, with all measured air‐kerma rates within a 2% range relative to the reference for a specific beam. However, some drift in response was observed, particularly for the RTI Piranha.
Conclusions
Both XMMs met the performance standards provided by their respective manufacturers. An additional calibration would result in increased measurement accuracy for some beam series. One limitation of this study is that the detectors were not subject to more rigorous clinical conditions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Zitieren
APA
Stasko, J. T., & Culberson, W. S. (2025). X‐ray multimeter performance under calibration laboratory conditions. Medical Physics, 52(9). https://doi.org/10.1002/mp.18106
Vancouver
Stasko JT, Culberson WS. X‐ray multimeter performance under calibration laboratory conditions. Medical Physics. 2025;52(9). doi:10.1002/mp.18106
BibTeX
@article{Stasko2025,
author = {Stasko, John T and Culberson, Wesley S},
title = {X‐ray multimeter performance under calibration laboratory conditions},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {9},
doi = {10.1002/mp.18106}
}
EUVIMED