Magnetic Resonance Imaging monitoring of histotripsy effects in agar phantom
Medical Physics · 2025-08 · 52 (8)
DOI: 10.1002/mp.18054 ↗At a glance
Journal article in Medical Physics, published online on 2025-08-13. Written by Anastasia Antoniou and 8 co-authors. Institution: Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Major MEDLINE topics: Agar, Phantoms, Imaging, Magnetic Resonance Imaging, High-Intensity Focused Ultrasound Ablation. Cited 2 times so far (Crossref); 50 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: Recovery and Resilience Facility, BRAINSONIC, Research and Innovation Foundation of Cyprus.
Abstract (original language)
Background
While the technical aspects of histotripsy have seen significant advancements, research on Magnetic Resonance Imaging (MRI) monitoring remains limited.
Purpose
This preliminary study explored the use of conventional T2‐Weighted (T2‐W) Turbo Spin Echo (TSE) imaging to monitor histotripsy lesions in a pure agar gel, as a potential research tool for MRI‐guided histotripsy (MRgHt).
Methods
Histotripsy experiments were conducted in a 3T MRI scanner using a 2% weight per volume agar solution in water. Pulsed Focused Ultrasound (FUS) was applied at a duty factor of 2%, with a pulse repetition period of 1s, delivering up to 2000 pulses. T2‐W TSE imaging was employed for intra‐procedural monitoring with 10s updates and post‐sonication assessment, with the first prioritizing rapid acquisition and the second focusing on resolution. Supplementary, MR thermometry tracked temperature variations throughout the procedure.
Results
Dynamic T2‐W TSE imaging provided insights into the progress of phantom disruption. Mild thermal effects served as early signs of phantom response, as mechanical effects required more time to manifest. Typical histotripsy lesion characteristics were observed, including a hyperintense core with an elongated shape and a thin hypointense border. The developed temperatures remained well below hyperthermia levels, reaching a maximum of 34°C.
Conclusions
The presented methodology, utilizing a pure agar gel within a simple, versatile setup alongside conventional T2‐W TSE imaging, holds promise for advancing MRgHt research.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus.
- Diagnostic and Interventional Radiology Department, German Medical Institute, Limassol, Cyprus.
- Department of Electronics and Information Engineering, Hangzhou Dianzi University, Hangzhou, Zhejiang, China.
- 3DMI Research Group, Department of Medical Physics, School of Medicine, University of Patras, Rion-Patras, Greece.
Funding
- Recovery and Resilience Facility
- BRAINSONIC
- Research and Innovation Foundation of Cyprus
Cite
APA
Antoniou, A., Evripidou, N., Georgiou, L., Christofi, A., Zhao, J., Yu, L., Li, W., Kagadis, G. C., & Damianou, C. (2025). Magnetic Resonance Imaging monitoring of histotripsy effects in agar phantom. Medical Physics, 52(8). https://doi.org/10.1002/mp.18054
Vancouver
Antoniou A, Evripidou N, Georgiou L, Christofi A, Zhao J, Yu L, et al. Magnetic Resonance Imaging monitoring of histotripsy effects in agar phantom. Medical Physics. 2025;52(8). doi:10.1002/mp.18054
BibTeX
@article{Antoniou2025,
author = {Antoniou, Anastasia and Evripidou, Nikolas and Georgiou, Leonidas and Christofi, Antonis and Zhao, Jufeng and Yu, Liyang and Li, Wenjun and Kagadis, George C and Damianou, Christakis},
title = {Magnetic Resonance Imaging monitoring of histotripsy effects in agar phantom},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {8},
doi = {10.1002/mp.18054}
}
EUVIMED