Hybrid phantom for lung CT: Design and validation

Paulo Roberto Costa, Gisell Ruiz Boiset, Elsa Bifano Pimenta, Raphael Moratta Vieira Rocha, Raissa Aline Santos Moura, Wagner Henrique Marques, Luuk J Oostveen, Bram Geurts, Marcio Valente Yamada Sawamura, Denise Yanikian Nersissian, Elisabeth Mateus Yoshimura, Ioannis Sechopoulos

Medical Physics · 2025-08 · 52 (8)

DOI: 10.1002/mp.17990 ↗

At a glance

Validierungsstudie in Medical Physics, published online on 2025-08-08. Written by Paulo Roberto Costa and 11 co-authors. Institution: Instituto de Física, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil. Major MEDLINE topics: Tomography, X-Ray Computed, Lung, Phantoms, Imaging, Lung Neoplasms. Cited 2 times so far (Crossref); 49 references. The full text is freely available (open access, CC BY 4.0). Funded by: Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP), Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq).

SantraukaSubjectsBibliografiniai duomenysCiteRelated studies

Abstract (original language)

Background

CT lung imaging protocols need to be optimized. This claim is especially important due to the possible introduction of low‐dose CT (LDCT) for lung cancer screening. Given the incorporation of non‐linear reconstructions and post‐processing, the use of phantoms that consider task‐based evaluation is needed. This is also true for acceptance and continuous QC use.

Purpose

To present and validate a lung‐CT hybrid phantom composed of two setups, one for task‐based image quality metrics and the other anthropomorphic.

Methods

The task‐based metrics setup was based on the well‐known Mercury phantom and the anthropomorphic setup named Freddie (from F igure of Me r it P e rformance evaluation of D etectability in D iagnostic CT I maging E quipment) was designed with the same basic dimensions of the Mercury phantom, but including pieces and materials for mimicking chest structures, such as tracheobronchial tree and lung parenchyma. This setup allows the inclusion of pieces of different sizes to mimic ground‐glass opacities, and sub‐solid and solid lung nodules. The validation of the phantom adopted three methods: comparative evaluation of the attenuation properties and the corresponding Hounsfield Units (HU) values of the selected materials; image assessment according to five chest radiologists and eight non‐radiologists’ observations (reader study), and measurement of task‐based metrics. Images of both setups were acquired using two clinical thorax protocols, both using automatic tube current modulation (TCM). Two x‐ray filter combinations were adopted. The images were reconstructed using a deep learning‐based algorithm.

Results

The agreement of nominal and observed HU values in the task‐based setup was within 15%, except for three (TangoBlack+, VeroClear, and HIPS) of the materials employed in the phantom construction, at some beam energies. In the reader study, synthetic solid nodules printed in VeroClear received average Likert scores 4.0 (range 3.0–4.0) from radiologists and 3 (range 2.6–3.8) from non‐radiologists, printed in TangoBlack+ received an average Likert score of 3.9 (range 3.8–4.2) from radiologists and 4.0 (range 3.8–4.4) from non‐radiologists, while those printed in HIPS scored an average Likert of 3.8 (range 3.3–3.9) from radiologists and 3.3 (range 3.1–3.3) from non‐radiologists. The synthetic ground‐glass opacities (GGO) nodules manufactured in EVA received an average Likert score of 3.8 (range 2.8–4.6) from radiologists and 4.3 (range 3.6–4.8) from non‐radiologists. The task‐based setup demonstrated detectability index variations across protocols influenced by the dose levels, voltage, and x‐ray beam filtration used.

Conclusions

The novelty of the proposed design is concentrated on the possibility of associating the response of the task‐based setup (Mercury) with a patient‐based setup (Freddie) in a unique phantom. This hybrid design enhances the potential to apply the detectability index for optimizing CT protocols in clinical scenarios.

Subjects

MeSH terms (MEDLINE)

Tomography, X-Ray ComputedLungPhantoms, ImagingLung NeoplasmsEarly Detection of CancerHumansEquipment Design

Author keywords

3D printingLDCTanthropomorphic phantomlung imaging

Authors and institutions

  • Instituto de Física, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil.
  • Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherland.
  • Departamento de Radiologia e Oncologia, Faculdade de Medicina, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil.
  • Hospital das Clinicas HCFMUSP, Faculdade de Medicina, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil.

Funding

  • Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
  • Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)

Cite

APA

Costa, P. R., Boiset, G. R., Pimenta, E. B., Rocha, R. M. V., Moura, R. A. S., Marques, W. H., Oostveen, L. J., Geurts, B., Sawamura, M. V. Y., Nersissian, D. Y., Yoshimura, E. M., & Sechopoulos, I. (2025). Hybrid phantom for lung CT: Design and validation. Medical Physics, 52(8). https://doi.org/10.1002/mp.17990

Vancouver

Costa PR, Boiset GR, Pimenta EB, Rocha RMV, Moura RAS, Marques WH, et al. Hybrid phantom for lung CT: Design and validation. Medical Physics. 2025;52(8). doi:10.1002/mp.17990

BibTeX
@article{Costa2025,
  author = {Costa, Paulo Roberto and Boiset, Gisell Ruiz and Pimenta, Elsa Bifano and Rocha, Raphael Moratta Vieira and Moura, Raissa Aline Santos and Marques, Wagner Henrique and Oostveen, Luuk J and Geurts, Bram and Sawamura, Marcio Valente Yamada and Nersissian, Denise Yanikian and Yoshimura, Elisabeth Mateus and Sechopoulos, Ioannis},
  title = {Hybrid phantom for lung CT: Design and validation},
  journal = {Medical Physics},
  year = {2025},
  volume = {52},
  number = {8},
  doi = {10.1002/mp.17990}
}
Start for free →