Quantitative CT reconstruction kernel harmonization for multi‐site lung cancer screening
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70325 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von David W Jordan und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University Hospitals Cleveland Medical Center, Case Western Reserve University School of Medicine, VA Northeast Ohio Healthc…. Hauptthemen laut MEDLINE: Lung Neoplasms, Phantoms, Imaging, Tomography, X-Ray Computed, Image Processing, Computer-Assisted. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Published reference CT protocols for lung cancer screening are not optimized to produce uniform image appearance across different scanner manufacturers and models or to conform to quantitative imaging profiles for robust small lung nodule size and volume measurements, which are important in clinical management of screen‐detected nodules.
Purpose
This study used widely available phantoms and software to identify lung cancer screening CT reconstructions that enable accurate and reproducible nodule size measurements and to match reconstructions across scanner manufacturers and models to provide a consistent image appearance to interpreting physicians.
Methods
ACR CT accreditation phantom scans were used to measure the modulation transfer function (MTF) and noise power spectrum (NPS) for various reconstruction kernels for six CT scanner models from three manufacturers. A reference kernel was chosen, and other kernels were matched based on agreement between MTF and NPS for each candidate kernel. Kernels were validated for conformance to the Quantitative Imaging Biomarkers Alliance (QIBA) Small Lung Nodule profile using a commercial phantom and conformance testing software.
Results
Medium sharp kernels with similar MTF and NPS curves and QIBA‐compliant imaging performance were identified for the six scanner models. One medium smooth kernel with iterative reconstruction and one common lung kernel did not conform to the QIBA profile.
Conclusions
MTF and NPS comparisons to a validated reference CT protocol can be used to identify candidate scanner‐specific reconstructions appropriate for lung cancer screening that conform to the QIBA protocol, thus producing accurate and reproducible nodule size measurements, and that should provide uniform image appearance to the interpreting radiologist.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, University Hospitals Cleveland Medical Center, Case Western Reserve University School of Medicine, VA Northeast Ohio Healthcare System, Cleveland, Ohio, USA.
- Radiation Safety, VA Northeast Ohio Healthcare System, Cleveland, Ohio, USA.
ORCID-Profile: David W Jordan
Zitieren
APA
Jordan, D. W., & Misseldine, R. E. (2025). Quantitative CT reconstruction kernel harmonization for multi‐site lung cancer screening. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70325
Vancouver
Jordan DW, Misseldine RE. Quantitative CT reconstruction kernel harmonization for multi‐site lung cancer screening. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70325
BibTeX
@article{Jordan2025,
author = {Jordan, David W and Misseldine, Ryan E},
title = {Quantitative CT reconstruction kernel harmonization for multi‐site lung cancer screening},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70325}
}
EUVIMED