Hybrid phantom for lung CT: Design and validation
Medical Physics · 2025-08 · 52 (8)
DOI: 10.1002/mp.17990 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Medical Physics, online erschienen am 08.08.2025. Verfasst von Paulo Roberto Costa und 11 weiteren Autor:innen. Beteiligte Einrichtung: Instituto de Física, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Lung, Phantoms, Imaging, Lung Neoplasms. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP), Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq).
Abstract (Originalsprache)
Background
CT lung imaging protocols need to be optimized. This claim is especially important due to the possible introduction of low‐dose CT (LDCT) for lung cancer screening. Given the incorporation of non‐linear reconstructions and post‐processing, the use of phantoms that consider task‐based evaluation is needed. This is also true for acceptance and continuous QC use.
Purpose
To present and validate a lung‐CT hybrid phantom composed of two setups, one for task‐based image quality metrics and the other anthropomorphic.
Methods
The task‐based metrics setup was based on the well‐known Mercury phantom and the anthropomorphic setup named Freddie (from F igure of Me r it P e rformance evaluation of D etectability in D iagnostic CT I maging E quipment) was designed with the same basic dimensions of the Mercury phantom, but including pieces and materials for mimicking chest structures, such as tracheobronchial tree and lung parenchyma. This setup allows the inclusion of pieces of different sizes to mimic ground‐glass opacities, and sub‐solid and solid lung nodules. The validation of the phantom adopted three methods: comparative evaluation of the attenuation properties and the corresponding Hounsfield Units (HU) values of the selected materials; image assessment according to five chest radiologists and eight non‐radiologists’ observations (reader study), and measurement of task‐based metrics. Images of both setups were acquired using two clinical thorax protocols, both using automatic tube current modulation (TCM). Two x‐ray filter combinations were adopted. The images were reconstructed using a deep learning‐based algorithm.
Results
The agreement of nominal and observed HU values in the task‐based setup was within 15%, except for three (TangoBlack+, VeroClear, and HIPS) of the materials employed in the phantom construction, at some beam energies. In the reader study, synthetic solid nodules printed in VeroClear received average Likert scores 4.0 (range 3.0–4.0) from radiologists and 3 (range 2.6–3.8) from non‐radiologists, printed in TangoBlack+ received an average Likert score of 3.9 (range 3.8–4.2) from radiologists and 4.0 (range 3.8–4.4) from non‐radiologists, while those printed in HIPS scored an average Likert of 3.8 (range 3.3–3.9) from radiologists and 3.3 (range 3.1–3.3) from non‐radiologists. The synthetic ground‐glass opacities (GGO) nodules manufactured in EVA received an average Likert score of 3.8 (range 2.8–4.6) from radiologists and 4.3 (range 3.6–4.8) from non‐radiologists. The task‐based setup demonstrated detectability index variations across protocols influenced by the dose levels, voltage, and x‐ray beam filtration used.
Conclusions
The novelty of the proposed design is concentrated on the possibility of associating the response of the task‐based setup (Mercury) with a patient‐based setup (Freddie) in a unique phantom. This hybrid design enhances the potential to apply the detectability index for optimizing CT protocols in clinical scenarios.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Instituto de Física, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil.
- Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherland.
- Departamento de Radiologia e Oncologia, Faculdade de Medicina, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil.
- Hospital das Clinicas HCFMUSP, Faculdade de Medicina, Universidade de São Paulo (USP), São Paulo, São Paulo, Brazil.
Förderung
- Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
- Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
Zitieren
APA
Costa, P. R., Boiset, G. R., Pimenta, E. B., Rocha, R. M. V., Moura, R. A. S., Marques, W. H., Oostveen, L. J., Geurts, B., Sawamura, M. V. Y., Nersissian, D. Y., Yoshimura, E. M., & Sechopoulos, I. (2025). Hybrid phantom for lung CT: Design and validation. Medical Physics, 52(8). https://doi.org/10.1002/mp.17990
Vancouver
Costa PR, Boiset GR, Pimenta EB, Rocha RMV, Moura RAS, Marques WH, et al. Hybrid phantom for lung CT: Design and validation. Medical Physics. 2025;52(8). doi:10.1002/mp.17990
BibTeX
@article{Costa2025,
author = {Costa, Paulo Roberto and Boiset, Gisell Ruiz and Pimenta, Elsa Bifano and Rocha, Raphael Moratta Vieira and Moura, Raissa Aline Santos and Marques, Wagner Henrique and Oostveen, Luuk J and Geurts, Bram and Sawamura, Marcio Valente Yamada and Nersissian, Denise Yanikian and Yoshimura, Elisabeth Mateus and Sechopoulos, Ioannis},
title = {Hybrid phantom for lung CT: Design and validation},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {8},
doi = {10.1002/mp.17990}
}
EUVIMED