Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron

Yao Hao, Tuyen Mai, Winter Green, Kelly McElvain, Erik Traneus, Sarah Patch

Medical Physics · 2025

DOI: 10.1002/mp.17954 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Yao Hao and 5 co-authors. Institution: Department of Radiation Oncology, Washington University in St Louis, St Louis, Missouri, USA. Major MEDLINE topics: Acoustics, Proton Therapy, Radiation Dosage. Cited 1 times so far (Crossref); 47 references. The full text is freely available in PubMed Central. Funded by: NCI NIH HHS, NIH SBIR.

SamenvattingSubjectsBibliografische gegevensCiteRelated studies

Abstract (original language)

Background

Minimization of range uncertainties is critical for precise and effective proton therapy. Thermoacoustic range verification is a promising non‐invasive technique for pinpointing the Bragg peak location in proton therapy.

Purpose

Verification of a prototype device for thermoacoustic range verification with minimal disruption to a clinical workflow for proton therapy delivered by a synchrocyclotron. Materials/Methods 5–8 pC/pulse were delivered to a gelatin ultrasound quality‐assurance phantom by a Mevion Hyperscan S250i system in 200–300 pulses/spot. Spots were laterally separated by 25 mm; energy layers were separated in depth by 10 MeV. Thermoacoustic and nuclear emissions were captured by a wireless thermoacoustic range verifier, with a radiation detector on board. Dose maps for individual beamlets were exported from RayStation 12A and used to compute initial values for finite element acoustic simulation, assuming Gaussian beam current with σ = 0.95 µs. The plan was re‐delivered with varying thicknesses of solid water equivalent phantom by Sun Nuclear placed on the snout. Range shifts and total dose were recomputed and loaded into RayStation for comparison to the original plan. To assess applicability to clinical plans, analysis was performed by averaging N = 16 pulses for which the radiation detector signal exceeded 50% of its maximum. Finite element results were bandpass filtered.

Results

Using only N = 16 thermoacoustic pulses yielded best‐ and worst‐case errors of and , averaged over all spots. Averaged over all spots only in a single layer, best‐ and worst‐case errors were 0.01.3 mm and −1.31.6 mm.

Conclusions

A wireless range verification research device that can be setup in less than 10 min provided millimeter range accuracy in an ultrasound quality assurance phantom.

Subjects

MeSH terms (MEDLINE)

AcousticsProton TherapyRadiation DosagePhantoms, ImagingRadiotherapy Dosage

Author keywords

compact‐wireless deviceparticle beamsthermoacoustic range verification

Authors and institutions

  • Department of Radiation Oncology, Washington University in St Louis, St Louis, Missouri, USA.
  • Acoustic Range Estimates, Chicago, Illinois, USA.
  • School of Medicine and Public Health, Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.
  • RaySearch Laboratories AB, Stockholm, Sweden.

Funding

  • NCI NIH HHS
  • NIH SBIR

Cite

APA

Hao, Y., Mai, T., Green, W., McElvain, K., Traneus, E., & Patch, S. (2025). Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron. Medical Physics. https://doi.org/10.1002/mp.17954

Vancouver

Hao Y, Mai T, Green W, McElvain K, Traneus E, Patch S. Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron. Medical Physics. 2025. doi:10.1002/mp.17954

BibTeX
@article{Hao2025,
  author = {Hao, Yao and Mai, Tuyen and Green, Winter and McElvain, Kelly and Traneus, Erik and Patch, Sarah},
  title = {Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17954}
}
Start for free →