Technical assessment of a novel vertical CT system for upright radiotherapy simulation and treatment planning
Medical Physics · 2026
DOI: 10.1002/mp.70312 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Jordan M Slagowski und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Human Oncology, University of Wisconsin - Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Radiotherapy Planning, Computer-Assisted, Patient Positioning. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NHLBI NIH HHS, National Cancer Institute of the National Institutes of Health.
Abstract (Originalsprache)
Background
Upright patient positioning may provide anatomical advantages and more cost‐effective proton therapy using a fixed beamline.
Purpose
To characterize image quality, imaging dose, and dose calculation accuracy for an upright CT scanner with a six‐degree‐of‐freedom patient positioning system.
Methods
CT imaging dose (CTDI vol ) was measured at 120 kVp for head and thorax protocols. Image quality was evaluated using an ACR‐464 phantom. Mean CT number accuracy was assessed within inserts of known material, and uniformity as the difference in values at the center and periphery of uniform phantoms. High‐contrast resolution was assessed by visible line pairs and modulation transfer function (MTF). Low‐contrast performance was quantified by contrast‐to‐noise‐ratio (CNR). Spatial integrity was evaluated between fiducials 100 mm apart. Hounsfield unit to mass density and stopping‐power‐ratio calibrations were performed. Proton and photon plans were optimized on upright CT scans of a thorax phantom in heterogeneous and homogeneous regions. Dose was forward computed on a registered recumbent CT scan and agreement evaluated using 3D gamma analysis.
Results
CTDI vol was 23.5 ± 0.02 mGy for the 16 cm head and 10.1 ± 0.01 mGy for the 32 cm body phantoms. Mean CT numbers (HU) were within the expected range for water (1.7) and acrylic (120.8). CT numbers were slightly [5–27 HU] out‐of‐range for air (−950.4), polyethylene (−78.8), and bone (823.0). Image uniformity was 20.2 HU and 35.0 HU for 20 and 48 cm diameter phantoms, respectively. Eight high‐contrast line pairs were visualized. The MTF equaled 4.4 cm −1 at 50% and 7.1 cm −1 at 10%. The median CNR was 0.93, below the ≥1.0 tolerance. Spatial integrity was ≤0.36 mm. Gamma pass rates were ≥99.8% for photon and ≥90.6% for proton plans with 1%/1 mm criteria, and ≥98.0% for all plans with 3%/2 mm criteria.
Conclusion
Upright CT image quality and dose calculation accuracy are acceptable for photon and proton radiotherapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Human Oncology, University of Wisconsin - Madison, Madison, Wisconsin, USA.
- Department of Medical Physics, University of Wisconsin - Madison, Madison, Wisconsin, USA.
- Leo Cancer Care, Inc., Middleton, Wisconsin, USA.
Förderung
- NHLBI NIH HHS
- National Cancer Institute of the National Institutes of Health
Zitieren
APA
Slagowski, J. M., Yan, Y., Miller, J. R., Hayes, J. W., Hoffman, C. A., Kang, M., & Glide-Hurst, C. K. (2026). Technical assessment of a novel vertical CT system for upright radiotherapy simulation and treatment planning. Medical Physics. https://doi.org/10.1002/mp.70312
Vancouver
Slagowski JM, Yan Y, Miller JR, Hayes JW, Hoffman CA, Kang M, et al. Technical assessment of a novel vertical CT system for upright radiotherapy simulation and treatment planning. Medical Physics. 2026. doi:10.1002/mp.70312
BibTeX
@article{Slagowski2026,
author = {Slagowski, Jordan M and Yan, Yuhao and Miller, Jessica R and Hayes, John W and Hoffman, Carson A and Kang, Minglei and Glide-Hurst, Carri K},
title = {Technical assessment of a novel vertical CT system for upright radiotherapy simulation and treatment planning},
journal = {Medical Physics},
year = {2026},
doi = {10.1002/mp.70312}
}
EUVIMED