Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron
Medical Physics · 2025
DOI: 10.1002/mp.17954 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Yao Hao und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Washington University in St Louis, St Louis, Missouri, USA. Hauptthemen laut MEDLINE: Acoustics, Proton Therapy, Radiation Dosage. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIH SBIR.
Abstract (Originalsprache)
Background
Minimization of range uncertainties is critical for precise and effective proton therapy. Thermoacoustic range verification is a promising non‐invasive technique for pinpointing the Bragg peak location in proton therapy.
Purpose
Verification of a prototype device for thermoacoustic range verification with minimal disruption to a clinical workflow for proton therapy delivered by a synchrocyclotron. Materials/Methods 5–8 pC/pulse were delivered to a gelatin ultrasound quality‐assurance phantom by a Mevion Hyperscan S250i system in 200–300 pulses/spot. Spots were laterally separated by 25 mm; energy layers were separated in depth by 10 MeV. Thermoacoustic and nuclear emissions were captured by a wireless thermoacoustic range verifier, with a radiation detector on board. Dose maps for individual beamlets were exported from RayStation 12A and used to compute initial values for finite element acoustic simulation, assuming Gaussian beam current with σ = 0.95 µs. The plan was re‐delivered with varying thicknesses of solid water equivalent phantom by Sun Nuclear placed on the snout. Range shifts and total dose were recomputed and loaded into RayStation for comparison to the original plan. To assess applicability to clinical plans, analysis was performed by averaging N = 16 pulses for which the radiation detector signal exceeded 50% of its maximum. Finite element results were bandpass filtered.
Results
Using only N = 16 thermoacoustic pulses yielded best‐ and worst‐case errors of and , averaged over all spots. Averaged over all spots only in a single layer, best‐ and worst‐case errors were 0.01.3 mm and −1.31.6 mm.
Conclusions
A wireless range verification research device that can be setup in less than 10 min provided millimeter range accuracy in an ultrasound quality assurance phantom.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Washington University in St Louis, St Louis, Missouri, USA.
- Acoustic Range Estimates, Chicago, Illinois, USA.
- School of Medicine and Public Health, Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.
- RaySearch Laboratories AB, Stockholm, Sweden.
Förderung
- NCI NIH HHS
- NIH SBIR
Zitieren
APA
Hao, Y., Mai, T., Green, W., McElvain, K., Traneus, E., & Patch, S. (2025). Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron. Medical Physics. https://doi.org/10.1002/mp.17954
Vancouver
Hao Y, Mai T, Green W, McElvain K, Traneus E, Patch S. Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron. Medical Physics. 2025. doi:10.1002/mp.17954
BibTeX
@article{Hao2025,
author = {Hao, Yao and Mai, Tuyen and Green, Winter and McElvain, Kelly and Traneus, Erik and Patch, Sarah},
title = {Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17954}
}
EUVIMED