Development and experimental validation of an in‐house treatment planning system with greedy energy layer optimization for fast IMPT
Medical Physics · 2025
DOI: 10.1002/mp.17941 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Medical Physics, 2025. Verfasst von Aoxiang Wang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, China. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Proton Therapy. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Background
Intensity‐modulated proton therapy (IMPT) using pencil beam technique scans tumor in a layer by layer, then spot by spot manner. It can provide highly conformal dose to tumor targets and spare nearby organs‐at‐risk (OAR). Fast delivery of IMPT can improve patient comfort and reduce motion‐induced uncertainties. Since energy layer switching time dominants the plan delivery time, reducing the number of energy layers is important for improving delivery efficiency. Although various energy layer optimization (ELO) methods exist, they are rarely experimentally validated or clinically implemented, since it is technically challenging to integrate these methods into commercially available treatment planning system (TPS) that is not open‐source.
Purpose
This work develops and experimentally validates an in‐house TPS (IH‐TPS) that incorporates a novel ELO method for the purpose of fast IMPT.
Methods
The dose calculation accuracy of IH‐TPS is verified against the measured beam data and the RayStation TPS. For treatment planning, a novel ELO method via greed selection algorithm is proposed to reduce energy layer switching time and total plan delivery time. To validate the planning accuracy of IH‐TPS, the 3D gamma index is calculated between IH‐TPS plans and RayStation plans for various scenarios. Patient‐specific quality‐assurance (QA) verifications are conducted to experimentally verify the delivered dose from the IH‐TPS plans for several clinical cases.
Results
Dose distributions in IH‐TPS matched with those from RayStation TPS, with 3D gamma index results exceeding 95% (2 mm, 2%). The ELO method significantly reduced the delivery time while maintaining plan quality. For instance, in a brain case, the number of energy layers was reduced from 78 to 40 (reduction of 38 layers), leading to a 62% reduction in total delivery time. Patient‐specific QA validation with the IBA ProteusONE proton machine confirmed a > 95% pass rate for all cases.
Conclusions
An IH‐TPS equipped with a novel ELO algorithm is developed and experimentally validated for the purpose of fast IMPT, with enhanced delivery efficiency and preserved plan quality.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, China.
- Department of Radiation Oncology, University of Kansas Medical Center, Kansas, USA.
Förderung
- NCI NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Wang, A., Zhu, Y. N., Setianegara, J., Lin, Y., Xiao, P., Xie, Q., & Gao, H. (2025). Development and experimental validation of an in‐house treatment planning system with greedy energy layer optimization for fast IMPT. Medical Physics. https://doi.org/10.1002/mp.17941
Vancouver
Wang A, Zhu YN, Setianegara J, Lin Y, Xiao P, Xie Q, et al. Development and experimental validation of an in‐house treatment planning system with greedy energy layer optimization for fast IMPT. Medical Physics. 2025. doi:10.1002/mp.17941
BibTeX
@article{Wang2025,
author = {Wang, Aoxiang and Zhu, Ya-Nan and Setianegara, Jufri and Lin, Yuting and Xiao, Peng and Xie, Qingguo and Gao, Hao},
title = {Development and experimental validation of an in‐house treatment planning system with greedy energy layer optimization for fast IMPT},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17941}
}
EUVIMED