Reference datasets for commissioning of model‐based dose calculation algorithms for electronic brachytherapy

Iymad R Mansour, Christian Valdes-Cortez, David Santiago Ayala Alvarez, Francisco Berumen, Jean-Simon Côte, Gaël Ndoutoume-Paquet, Peter G F Watson, Jan Seuntjens, Facundo Ballester, Ernesto Mainegra-Hing, Rowan M Thomson, Luc Beaulieu und 1 weitere

Medical Physics · 2025-07 · 52 (7)

DOI: 10.1002/mp.17872 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 25.05.2025. Verfasst von Iymad R Mansour und 12 weiteren Autor:innen. Beteiligte Einrichtung: Carleton Laboratory for Radiotherapy Physics, Department of Physics, Carleton University, Ottawa, Canada. Hauptthemen laut MEDLINE: Brachytherapy, Algorithms, Radiotherapy Planning, Computer-Assisted, Radiation Dosage. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Conselleria de Educación, Universidades y Empleo (GVA), AEI, "ERDF A way of making Europe" by the "European Union NextGenerationEU/PRTR".

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose This work provides the first two clinical test cases for commissioning electronic brachytherapy (eBT) model‐based dose calculation algorithms (MBDCAs) for skin irradiation using surface applicators. Acquisition and Validation Methods The test cases utilize the INTRABEAM 30 mm surface applicator. Test Case I : water phantom is used to evaluate the algorithm's performance in a uniform medium consisting of a voxelized water cube surrounded by air. Test Case II: Surface eBT represents a heterogeneous medium with four distinct layers: skin tissue, adipose tissue, cortical bone, and soft tissue. Treatment plans for both cases were created and exported into the Radiance treatment planning system (TPS). Dose‐to‐medium calculations were then performed using this Monte Carlo (MC)‐based TPS and compared with MC simulations conducted independently by three different groups using two codes: EGSnrc and PENELOPE. The results agreed within expected Type A and B statistical uncertainties. Data Format and Usage Notes The dataset is available online at https://doi.org/10.52519/00005. A proprietary file designed for use within Radiance containing CT images and the treatment plan for both test cases, the LINAC modeling, and the CT calibration are included, as well as reference MC and TPS dose data in RTdose format and all files required to run the MC simulations. Potential Applications This dataset serves as a valuable resource for commissioning eBT MBDCAs and lays the groundwork for developing clinical test cases for other eBT systems. It is also a helpful educational tool for exploring various eBT devices and their advantages and drawbacks. Furthermore, brachytherapy researchers seeking a benchmark for dosimetric calculations in the low‐energy domain will find this dataset indispensable.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyAlgorithmsRadiotherapy Planning, Computer-AssistedRadiation DosageDatasets as TopicRadiotherapy DosageMonte Carlo MethodHumansPhantoms, Imaging

Stichwörter der Autor:innen

Monte Carlo (MC)intraoperative radiation therapy (IORT)model based dose calculation algorithms (MBDCA)

Autor:innen und Institutionen

  • Carleton Laboratory for Radiotherapy Physics, Department of Physics, Carleton University, Ottawa, Canada.
  • Princess Margaret Cancer Centre, Radiation Medicine Program, University Health Network, Toronto, Canada.
  • Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada.
  • Nuclear Medicine Department, Hospital Regional de Antofagasta, Antofagasta, Chile.
  • Department of Physics and Medical Physics Unit, McGill University, Montreal, Canada.
  • Service de Physique Médicale et de Radioprotection, Centre Intégré de Cancérologie, CHU de Québec- Université Laval et Centre de recherche du CHU de Québec, Québec, Canada.

Förderung

  • Conselleria de Educación, Universidades y Empleo (GVA)
  • AEI
  • "ERDF A way of making Europe" by the "European Union NextGenerationEU/PRTR"
  • JCyL
  • Fonds de Recherche du Québec-Santé
  • CIHR
  • Natural Sciences and Engineering Research Council
  • Agencia Estatal de Investigación

Zitieren

APA

Mansour, I. R., Valdes-Cortez, C., Ayala Alvarez, D. S., Berumen, F., Côte, J. S., Ndoutoume-Paquet, G., Watson, P. G. F., Seuntjens, J., Ballester, F., Mainegra-Hing, E., Thomson, R. M., Beaulieu, L., & Vijande, J. (2025). Reference datasets for commissioning of model‐based dose calculation algorithms for electronic brachytherapy. Medical Physics, 52(7). https://doi.org/10.1002/mp.17872

Vancouver

Mansour IR, Valdes-Cortez C, Ayala Alvarez DS, Berumen F, Côte JS, Ndoutoume-Paquet G, et al. Reference datasets for commissioning of model‐based dose calculation algorithms for electronic brachytherapy. Medical Physics. 2025;52(7). doi:10.1002/mp.17872

BibTeX
@article{Mansour2025,
  author = {Mansour, Iymad R and Valdes-Cortez, Christian and Ayala Alvarez, David Santiago and Berumen, Francisco and Côte, Jean-Simon and Ndoutoume-Paquet, Gaël and Watson, Peter G F and Seuntjens, Jan and Ballester, Facundo and Mainegra-Hing, Ernesto and Thomson, Rowan M and Beaulieu, Luc and Vijande, Javier},
  title = {Reference datasets for commissioning of model‐based dose calculation algorithms for electronic brachytherapy},
  journal = {Medical Physics},
  year = {2025},
  volume = {52},
  number = {7},
  doi = {10.1002/mp.17872}
}
Kostenlos starten →