Practical experience with initial quality assurance of a high dose rate brachytherapy grid‐based Boltzmann solver algorithm

Jessica M Fagerstrom iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14392 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Jessica M Fagerstrom. Beteiligte Einrichtung: Department of Radiation Oncology, University of Washington, Seattle, Washington, USA. Hauptthemen laut MEDLINE: Brachytherapy, Radiotherapy Dosage, Algorithms, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The purpose of this study was to validate the use of a model‐based dose calculation algorithm (MBDCA), Acuros BV, for high dose rate brachytherapy treatment planning for a community‐based hospital with a Bravos afterloader. Based on published AAPM recommendations, this work details a practical approach for community‐based clinics to complete initial validation of Acuros BV, in order to add a MBDCA to a TG‐43 based brachytherapy treatment planning program.

Methods

Source dimensions and materials used in Acuros BV and TG‐43 source models were compared to the physical source. TG‐186 testing was completed with standardized test cases externally calculated with Monte Carlo compared to locally calculated with Acuros BV. Point doses calculated using TG‐43 were compared to those calculated with Acuros BV in water at various dose grid settings. Secondary dose check software was used to evaluate dose distributions resembling clinical patient plans, both in water and on CT datasets representative of patient anatomy.

Results

The major source of discrepancy of source models was the length of modeled steel cable. TG‐186 testing showed that the largest differences between Monte Carlo and Acuros BV dose distributions were located along the source axis for cases calculated in water, as well as located in regions of high dose gradients and within the applicator for the case calculated with a generic shielded applicator. An audit of point doses calculated with both TG‐43 and Acuros BV in water found that dose grid settings significantly affected agreement. Secondary dose check software indicated that Acuros BV functioned satisfactorily, and a 5% threshold was adopted for secondary dose checks on gynecologic plans.

Conclusion

This validation process indicated that Acuros BV met expected standards and affirmed its suitability for integration into this clinical practice's brachytherapy treatment planning.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyRadiotherapy DosageAlgorithmsRadiotherapy Planning, Computer-AssistedMonte Carlo MethodQuality Assurance, Health CarePhantoms, ImagingHumansSoftwareNeoplasmsRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

HDRMBDCAbrachytherapycommissioningdose calculationquality assurancetreatment planning

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Washington, Seattle, Washington, USA.
  • Department of Radiation Oncology, Kaiser Permanente, Seattle, Washington, USA.

ORCID-Profile: Jessica M Fagerstrom

Zitieren

APA

Fagerstrom, J. M. (2024). Practical experience with initial quality assurance of a high dose rate brachytherapy grid‐based Boltzmann solver algorithm. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14392

Vancouver

Fagerstrom JM. Practical experience with initial quality assurance of a high dose rate brachytherapy grid‐based Boltzmann solver algorithm. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14392

BibTeX
@article{Fagerstrom2024,
  author = {Fagerstrom, Jessica M},
  title = {Practical experience with initial quality assurance of a high dose rate brachytherapy grid‐based Boltzmann solver algorithm},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14392}
}
Kostenlos starten →