Quantifying the dosimetric accuracy of expiration‐gated stereotactic lung radiotherapy

Daan Hoffmans, Isabel Remmerts de Vries, Max Dahele, Wilko Verbakel

Medical Physics · 2025

DOI: 10.1002/mp.17743 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Daan Hoffmans und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Amsterdam UMC location Vrije Universiteit Amsterdam, Amsterdam, The Netherlands. Hauptthemen laut MEDLINE: Radiosurgery, Lung Neoplasms, Exhalation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In stereotactic body radiotherapy, a form of motion management is often applied to mobile lung tumors. Gated radiotherapy is such form of motion management in which the radiation beam is switched on or off depending on the actual tumor position. Compared to inspiration, the tumor position is typically more stable during expiration. Also, the tumor spends more time in expiration position. Therefore, we often consider expiration‐gating for patients with relatively large tumor motion.

Purpose

We validated dosimetric accuracy of expiration‐gated stereotactic lung radiotherapy by means of phantom measurements and modeling the effects of residual motion in patients.

Methods

Dose profiles from film measurements in a respiratory‐motion phantom were compared to dose calculations, for different expiration gating methods. Fluoroscopic real‐time tumor tracking was used to produce a convolution kernel which was applied to the calculated dose distribution to model dosimetric effects of residual motion. This convolution method was validated against film measurements and then retrospectively applied to clinical tumor tracking data of five patients. In addition, clinical tumor motion data was manipulated to simulate the effect of a short breathing period of 2 s and prolonged gating latency of 500 ms.

Results

A good agreement between calculated and measured dose was found when amplitude gating was used (100% gamma pass rate, 3%/2 mm). For phase gating, good agreement required a stable breathing period. Measurements showed good performance of the convolution method (gamma pass rate > 99%). For the clinical data, we found a maximal dose shift of 2.4 mm, introduced by residual tumor motion or respiratory drift. For all patients, the size of the ITV was adequate to account for this dose shift. Simulating higher breathing speed in combination with large latency values resulted in dosimetric shifts that were larger than the PTV margin.

Conclusion

Amplitude gating is robust for irregular breathing patterns. Expiration‐gating is a dosimetrically accurate method of treatment delivery provided that during delivery there is a prompt reaction to respiratory drift and the latency of the gating system is short.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryLung NeoplasmsExhalationHumansPhantoms, ImagingRadiometryRadiotherapy DosageMovementRespiration

Stichwörter der Autor:innen

exhale gatingexpiration gatingresidual motiontumor tracking

Autor:innen und Institutionen

  • Department of Radiation Oncology, Amsterdam UMC location Vrije Universiteit Amsterdam, Amsterdam, The Netherlands.
  • Cancer Center Amsterdam, Cancer Treatment and Quality of Life, Amsterdam, The Netherlands.
  • Varian Medical Systems, Radiotherapy Solutions, Palo Alto, USA.

Zitieren

APA

Hoffmans, D., Remmerts de Vries, I., Dahele, M., & Verbakel, W. (2025). Quantifying the dosimetric accuracy of expiration‐gated stereotactic lung radiotherapy. Medical Physics. https://doi.org/10.1002/mp.17743

Vancouver

Hoffmans D, Remmerts de Vries I, Dahele M, Verbakel W. Quantifying the dosimetric accuracy of expiration‐gated stereotactic lung radiotherapy. Medical Physics. 2025. doi:10.1002/mp.17743

BibTeX
@article{Hoffmans2025,
  author = {Hoffmans, Daan and Remmerts de Vries, Isabel and Dahele, Max and Verbakel, Wilko},
  title = {Quantifying the dosimetric accuracy of expiration‐gated stereotactic lung radiotherapy},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17743}
}
Kostenlos starten →