Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Ying Zhang und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics and Biomedical Engineering, University College London, London, UK. Hauptthemen laut MEDLINE: Proton Therapy, Head and Neck Neoplasms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Anatomical changes present a major source of uncertainty in head and neck (H&N) cancer treatment. Accurate modeling of these changes is important for enhancing treatment precision and supporting better outcomes.
Purpose
The purpose of this study is to assess different anatomical uncertainty modeling methods in robust optimization for H&N cancer proton therapy.
Methods
This retrospective study involved five nasopharynx radiotherapy patients. We compared conventional robust optimization with anatomical robust optimization (aRO): (1) conventional robust optimization (cRO‐3 mm), which used 3 mm setup shift and 3% range uncertainty. (2) aRO_AM which used three predicted images from an AM capturing systematic anatomical changes, with a 1 mm setup shift and 3% range uncertainty. (3) aRO_PM, which used three predicted images from a probability model (PM) capturing the most probable deformations, also with a 1 mm setup shift and 3% range uncertainty. We assessed weekly dose coverage of the clinical target volumes (CTVs). Normal tissue complication probability (NTCP) for grade 2 xerostomia and grade 2 dysphagia were calculated using the accumulated nominal dose (without errors). Results aRO_PM outperformed cRO‐3 mm and aRO_AM, consistently achieving V94 95% for all cases across treatment weeks. Additionally, aRO_PM reduced the NTCP for grade 2 xerostomia by an average of 4.88 %, with a maximum reduction of 8.03%, and reduced the NTCP for grade 2 dysphagia by an average of 1.80%, with a maximum reduction of 4.23 %.
Conclusion
The PM demonstrates potential for improving robust optimization by effectively managing anatomical uncertainties in H&N cancer proton therapy, thereby enhancing treatment effectiveness.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics and Biomedical Engineering, University College London, London, UK.
- Department of Radiation Oncology, University Nebraska Medical Center, Omaha, USA.
ORCID-Profile: Mark Ka Heng Chan
Zitieren
APA
Zhang, Y., & Chan, M. K. H. (2025). Enhancing accuracy in proton therapy: The impact of geometric uncertainty models in head and neck cancer treatment. Medical Physics. https://doi.org/10.1002/mp.17698
Vancouver
Zhang Y, Chan MKH. Enhancing accuracy in proton therapy: The impact of geometric uncertainty models in head and neck cancer treatment. Medical Physics. 2025. doi:10.1002/mp.17698
BibTeX
@article{Zhang2025,
author = {Zhang, Ying and Chan, Mark Ka Heng},
title = {Enhancing accuracy in proton therapy: The impact of geometric uncertainty models in head and neck cancer treatment},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17698}
}
EUVIMED