Dosimetric study of synchrotron rapid beam off control and skip spot function for high beam intensity proton therapy

Masashi Yagi, Keith M Furutani, Toshiyuki Ogata, Takuya Nomura, Masumi Umezawa, Xiaoying Liang, Kei Yamada, Hideya Yamazaki, Shinichi Shimizu, Chris J Beltran

Medical Physics · 2025

DOI: 10.1002/mp.17589 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Masashi Yagi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Carbon Ion Radiotherapy, Osaka University Graduate School of Medicine, Osaka, Japan. Hauptthemen laut MEDLINE: Proton Therapy, Synchrotrons, Radiometry. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: JSPS KAKENHI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

All Hitachi proton pencil beam scanning facilities currently use discrete spot scanning (DSS). Mayo Clinic Florida (MCF) is installing a Hitachi particle therapy system with advanced technologies, including fast scan speeds, high beam intensity, rapid beam off control (RBOC), a skip spot function, and proton pencil beam scanning using dose driven continuous scanning (DDCS). A potential concern of RBOC is the generation of a shoulder at the end of the normal spot delivery due to a flap spot (FS) with a flap dose (FD), which has been investigated for carbon synchrotron but not for proton delivery. While investigated, for instance, for Hitachi's installation at MCF, this methodology could be applicable for all future high intensity proton deliveries.

Purpose

No Hitachi proton facility currently uses the proposed RBOC. This study aimed to understand the dosimetric impact of proton FD at MCF by simulating the FS with a Hitachi proton machine in research mode, reflecting the higher proton intensities expected with RBOC at MCF.

Method

Experiments were conducted to simulate MCF RBOC at Kyoto Prefecture University of Medicine (KPUM) in research mode, reducing delay time (Td) from 1.5 ms to 0.1 ms. 5,000 contiguous spots were delivered on the central axis for proton energies of 70.2, 142.5, and 220.0 MeV; at normal, high dose rate (HDR), and ultra‐high dose rate (uHDR) intensities; and at vertical and horizontal gantry angles for different Td. Measurements were taken using a fast oscilloscope and the nozzle's spot position monitor (SPM) and dose monitor (DM). A model was developed to predict FD dependence on beam intensity and assess the dosimetric impact for prostate and brain treatment plans. Two simulation types were planned: a flap DSS plan with FS at every spot and a flap DDCS plan with FS only at the end of each layer.

Result

FD was observed for RBOC with Td = 0.1 ms, showing no gantry angle dependence. FD increased with higher delayed dose rate (DDR), that is, beam intensity. The planning study showed dose volume histogram deterioration with increased FD compared to the clinical plan, but it was only significant for uHDR intensities. Deterioration was marginal in flap DSS plans for the HDR intensities planned at MCF, and flap DDCS plans were even less sensitive than flap DSS plans.

Conclusion

MCF is installing proton DDCS with higher beam intensities, a skip spot function, and fast beam‐off control. The resulting FD had an insignificant impact on dose distribution for two patient plans with both DSS and DDCS at the anticipated MCF intensities. However, significant dependence was observed in the case of uHDR. A method to measure the position and dose of the FS during commissioning is described in addition to recommendations for regular QA and log‐based proton patient‐specific quality assurance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapySynchrotronsRadiometryRadiotherapy DosageHumansTime Factors

Stichwörter der Autor:innen

dose driven continuous scanning (DDCS)flap doseproton therapy

Autor:innen und Institutionen

  • Department of Carbon Ion Radiotherapy, Osaka University Graduate School of Medicine, Osaka, Japan.
  • Division of Medical Physics, Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA.
  • Department of Radiology, Kyoto Prefectural University of Medicine, Kyoto, Japan.
  • Therapy System Business Division/Healthcare Business Group, Hitachi High-Tech Corporation, Chiba, Japan.

Förderung

  • JSPS KAKENHI

Zitieren

APA

Yagi, M., Furutani, K. M., Ogata, T., Nomura, T., Umezawa, M., Liang, X., Yamada, K., Yamazaki, H., Shimizu, S., & Beltran, C. J. (2025). Dosimetric study of synchrotron rapid beam off control and skip spot function for high beam intensity proton therapy. Medical Physics. https://doi.org/10.1002/mp.17589

Vancouver

Yagi M, Furutani KM, Ogata T, Nomura T, Umezawa M, Liang X, et al. Dosimetric study of synchrotron rapid beam off control and skip spot function for high beam intensity proton therapy. Medical Physics. 2025. doi:10.1002/mp.17589

BibTeX
@article{Yagi2025,
  author = {Yagi, Masashi and Furutani, Keith M and Ogata, Toshiyuki and Nomura, Takuya and Umezawa, Masumi and Liang, Xiaoying and Yamada, Kei and Yamazaki, Hideya and Shimizu, Shinichi and Beltran, Chris J},
  title = {Dosimetric study of synchrotron rapid beam off control and skip spot function for high beam intensity proton therapy},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17589}
}
Kostenlos starten →