Improved heterogeneity handling in the collapsed cone dose engine for brachytherapy

Freja Alpsten, Bob van Veelen, Christian Valdes-Cortez, Francisco Berumen, Anders Ahnesjö, Åsa Carlsson Tedgren

Medical Physics · 2025

DOI: 10.1002/mp.17434 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Freja Alpsten und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology-Pathology, Karolinska Institutet, Stockholm, Sweden. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Algorithms, Brachytherapy, Monte Carlo Method. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swedish Cancer Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Model‐based dose calculation algorithms (MBDCA), such as the Advanced Collapsed cone Engine (ACE) in Oncentra Brachy® can be used to overcome the limitations of the TG‐43 formalism. ACE is a point kernel superposition algorithm that calculates the total dose separated into primary, first‐scatter, and multiple‐scatter dose. Albeit ACE yields accurate results under most circumstances, several studies have reported underestimations of the dose to cortical bone. These underestimations are likely caused by approximations in the handling of multiple‐scatter dose for non‐water media. Such would result in noticeable deviations where the multiple‐scatter is a considerable fraction of the total dose, that is, at greater distances from the source.

Purpose

To improve and test the accuracy of the multiple‐scatter dose component in the ACE algorithm to remedy its inaccuracy for non‐water geometries.

Methods

A careful analysis of the transport and absorption of the multiple‐scatter energy fluence revealed an inconsistency in the scaling of energy absorption ratios for non‐water media of the multiple‐scatter kernel. We implemented an updated algorithm version, ACE corr , and tested it for three different geometries. All had a single 192 Ir‐source at the center of a cubic water phantom with a box‐shaped heterogeneity of either cortical bone or air, positioned at different distances from the source. Dose distributions for the three cases were calculated with ACE and ACE corr and compared to Monte Carlo simulations, using the percentage dose difference ratio as figure‐of‐merit. All dose calculation methods scored separately the dose deposited by primary, first‐scattered, and multiple‐scattered photons.

Results

The accuracy of the updated algorithm ACE corr was superior to ACE. In the cortical bone heterogeneity, the mean percentage dose difference ratio for the total dose improved from to (in the worst case) by our update. Less impact was seen in the air heterogeneity, where both ACE and ACE corr deviated less than 2% from the Monte Carlo results. The algorithm update mainly concerns the multiple‐scattered dose component, but an accompanying data processing update also had a small effect (0.5% difference) on the primary and first‐scattered dose. The calculation times were not affected.

Conclusions

The updates to ACE improved the accuracy of multiple‐scatter dose calculation for non‐water media, without increasing calculation times.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy DosageAlgorithmsBrachytherapyMonte Carlo MethodRadiotherapy Planning, Computer-AssistedPhantoms, ImagingScattering, RadiationHumansRadiation Dosage

Stichwörter der Autor:innen

brachytherapycollapsed conemodel‐based dose calculation algorithm

Autor:innen und Institutionen

  • Department of Oncology-Pathology, Karolinska Institutet, Stockholm, Sweden.
  • Department of Nuclear Medicin and Medical Physics, Karolinska University Hospital, Stockholm, Sweden.
  • Elekta Brachytherapy, Veenendaal, The Netherlands.
  • Nuclear Medicine Department, Hospital Regional de Antofagasta, Antofagasta, Chile.
  • Service de Physique Médicale et de Radioprotection, Centre Intégré de Cancérologie, CHU de Québec - Université Laval et Centre de recherche du CHU de Québec, Quebec, Canada.
  • Département de Physique, de Génie Physique et d'Optique et Centre de Recherche sur le Cancer, Université Laval, Quebec, Canada.

Förderung

  • Swedish Cancer Foundation

Zitieren

APA

Alpsten, F., van Veelen, B., Valdes-Cortez, C., Berumen, F., Ahnesjö, A., & Carlsson Tedgren, Å. (2025). Improved heterogeneity handling in the collapsed cone dose engine for brachytherapy. Medical Physics. https://doi.org/10.1002/mp.17434

Vancouver

Alpsten F, van Veelen B, Valdes-Cortez C, Berumen F, Ahnesjö A, Carlsson Tedgren Å. Improved heterogeneity handling in the collapsed cone dose engine for brachytherapy. Medical Physics. 2025. doi:10.1002/mp.17434

BibTeX
@article{Alpsten2025,
  author = {Alpsten, Freja and van Veelen, Bob and Valdes-Cortez, Christian and Berumen, Francisco and Ahnesjö, Anders and Carlsson Tedgren, Åsa},
  title = {Improved heterogeneity handling in the collapsed cone dose engine for brachytherapy},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17434}
}
Kostenlos starten →