Long‐term beam output stability of an accelerator‐based boron neutron capture therapy system

Shinya Komori iD, Akihiko Takeuchi iD, Ryohei Kato iD, Yuhei Yamazaki, Tomoaki Motoyanagi, Yuki Narita iD, Takahiro Kato iD, Yoshihiro Takai

Medical Physics · 2024

DOI: 10.1002/mp.17426 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Shinya Komori und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, Koriyama, Fukushima, Japan. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 48 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Accelerator‐based boron neutron capture therapy (AB‐BNCT) systems are becoming commercially available and are expected to be widely used in hospitals. To ensure the safety of BNCT, establishing a quality assurance (QA) program and properly managing the stability of the system are necessary. In particular, a high level of beam output stability is required to avoid accidents because beam output is a major factor in patient dose. However, no studies have analyzed the long‐term beam output stability of AB‐BNCT systems.

Purpose

This study aimed to retrospectively analyze the long‐term stability of the beam output by statistical process control (SPC) based on the QA results over 3 years.

Methods

The data analyzed are the results of daily QA (DQA) and weekly QA (WQA) in an AB‐BNCT system and were taken between June 2020 and September 2023. The evaluation of the stability of the beam output was based on the reaction rate between gold and neutrons calculated using the activation foil method using a gold foil. In DQA, which can be performed in a short time, the gold foil was applied directly to the beam irradiation aperture in air. In WQA, measurements were performed at the phantom surface, 2‐cm depth, and 6‐cm depth using a dedicated water phantom. The acquired data were retrospectively analyzed by individuals and a moving range chart (I‐MR chart), exponentially weighted moving average control chart (EWMA chart), and several process capability indexes (PCIs).

Results

Over 99% of the DQA I‐MR chart results were within control limits, whereas the WQA I‐MR chart results showed that 1.8%, 4.1%, and 2.0% of the measurements exceeded the control limits at the surface, 2‐cm depth, and 6‐cm depth, respectively. The variation in the reaction rate of the gold foil before and after the replacement of the target was 1.0; however, the C pmk of DQA and the 2‐ and 6‐cm depth WQAs between August 2021 and November 2022 in treatment course 2 were 0.83, 0.77, and 0.87, respectively, which were

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Boron Neutron Capture TherapyRetrospective StudiesParticle AcceleratorsTime FactorsQuality ControlHumansPhantoms, ImagingQuality Assurance, Health Care

Stichwörter der Autor:innen

acceleratorbeam outputboron neutron capture therapyquality assurancestatistical process control

Autor:innen und Institutionen

  • Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, Koriyama, Fukushima, Japan.
  • School of Health Sciences, Fukushima Medical University, Fukushima, Fukushima, Japan.
  • Department of Radiation Oncology, Southern Tohoku BNCT Research Center, Koriyama, Fukushima, Japan.

ORCID-Profile: Shinya Komori, Akihiko Takeuchi, Ryohei Kato, Yuki Narita, Takahiro Kato

Zitieren

APA

Komori, S., Takeuchi, A., Kato, R., Yamazaki, Y., Motoyanagi, T., Narita, Y., Kato, T., & Takai, Y. (2024). Long‐term beam output stability of an accelerator‐based boron neutron capture therapy system. Medical Physics. https://doi.org/10.1002/mp.17426

Vancouver

Komori S, Takeuchi A, Kato R, Yamazaki Y, Motoyanagi T, Narita Y, et al. Long‐term beam output stability of an accelerator‐based boron neutron capture therapy system. Medical Physics. 2024. doi:10.1002/mp.17426

BibTeX
@article{Komori2024,
  author = {Komori, Shinya and Takeuchi, Akihiko and Kato, Ryohei and Yamazaki, Yuhei and Motoyanagi, Tomoaki and Narita, Yuki and Kato, Takahiro and Takai, Yoshihiro},
  title = {Long‐term beam output stability of an accelerator‐based boron neutron capture therapy system},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17426}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche