First Monte Carlo beam model for ultra‐high dose rate radiotherapy with a compact electron LINAC
Medical Physics · 2024
DOI: 10.1002/mp.17031 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Tianyuan Dai und 14 weiteren Autor:innen. Beteiligte Einrichtung: Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA. Hauptthemen laut MEDLINE: Monte Carlo Method, Electrons, Radiotherapy Dosage, Particle Accelerators. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, National Natural Science Foundation of China, Shandong Provincial Natural Science Foundation.
Abstract (Originalsprache)
Background
FLASH radiotherapy based on ultra‐high dose rate (UHDR) is actively being studied by the radiotherapy community. Dedicated UHDR electron devices are currently a mainstay for FLASH studies.
Purpose
To present the first Monte Carlo (MC) electron beam model for the UHDR capable Mobetron (FLASH‐IQ) as a dose calculation and treatment planning platform for preclinical research and FLASH‐radiotherapy (RT) clinical trials.
Methods
The initial beamline geometry of the Mobetron was provided by the manufacturer, with the first‐principal implementation realized in the Geant4‐based GAMOS MC toolkit. The geometry and electron source characteristics, such as energy spectrum and beamline parameters, were tuned to match the central‐axis percentage depth dose (PDD) and lateral profiles for the pristine beam measured during machine commissioning. The thickness of the small foil in secondary scatter affected the beam model dominantly and was fine tuned to achieve the best agreement with commissioning data. Validation of the MC beam modeling was performed by comparing the calculated PDDs and profiles with EBT‐XD radiochromic film measurements for various combinations of applicators and inserts.
Results
The nominal 9 MeV electron FLASH beams were best represented by a Gaussian energy spectrum with mean energy of 9.9 MeV and variance (σ) of 0.2 MeV. Good agreement between the MC beam model and commissioning data were demonstrated with maximal discrepancy < 3% for PDDs and profiles. Hundred percent gamma pass rate was achieved for all PDDs and profiles with the criteria of 2 mm/3%. With the criteria of 2 mm/2%, maximum, minimum and mean gamma pass rates were (100.0%, 93.8%, 98.7%) for PDDs and (100.0%, 96.7%, 99.4%) for profiles, respectively.
Conclusions
A validated MC beam model for the UHDR capable Mobetron is presented for the first time. The MC model can be utilized for direct dose calculation or to generate beam modeling input required for treatment planning systems for FLASH‐RT planning. The beam model presented in this work should facilitate translational and clinical FLASH‐RT for trials conducted on the Mobetron FLASH‐IQ platform.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA.
- Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, Shandong, China.
- UT Southwestern Medical Center, Dallas, Texas, USA.
- IntraOp Medical Corporation, Sunnyvale, California, USA.
- Department of Medicine, Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire, USA.
- Dartmouth Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire, USA.
Förderung
- NCI NIH HHS
- National Natural Science Foundation of China
- Shandong Provincial Natural Science Foundation
- NIBIB NIH HHS
- National Cancer Institute
- Natural Science Foundation of Shandong Province
Zitieren
APA
Dai, T., Sloop, A. M., Rahman, M. R., Sunnerberg, J. P., Clark, M. A., Young, R., Adamczyk, S., Von Voigts-Rhetz, P., Patane, C., Turk, M., Jarvis, L., Pogue, B. W., Gladstone, D. J., Bruza, P., & Zhang, R. (2024). First Monte Carlo beam model for ultra‐high dose rate radiotherapy with a compact electron LINAC. Medical Physics. https://doi.org/10.1002/mp.17031
Vancouver
Dai T, Sloop AM, Rahman MR, Sunnerberg JP, Clark MA, Young R, et al. First Monte Carlo beam model for ultra‐high dose rate radiotherapy with a compact electron LINAC. Medical Physics. 2024. doi:10.1002/mp.17031
BibTeX
@article{Dai2024,
author = {Dai, Tianyuan and Sloop, Austin M and Rahman, Mahbubur R and Sunnerberg, Jacob P and Clark, Megan A and Young, Ralph and Adamczyk, Sebastian and Von Voigts-Rhetz, Philip and Patane, Chris and Turk, Michael and Jarvis, Lesley and Pogue, Brian W and Gladstone, David J and Bruza, Petr and Zhang, Rongxiao},
title = {First Monte Carlo beam model for ultra‐high dose rate radiotherapy with a compact electron LINAC},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17031}
}
EUVIMED