Technical note: Determination of CQ$C_{Q}$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities

Peter G F Watson, Stephen Davis, Wesley S Culberson

Medical Physics · 2024

DOI: 10.1002/mp.17345 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Peter G F Watson and 2 co-authors. Institution: Medical Physics Unit, McGill University, Montreal, Quebec, Canada. Major MEDLINE topics: Radiometry, Monte Carlo Method. Cited 1 times so far (Crossref); 13 references. The full text is freely available (open access, CC BY 4.0).

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

The formalism proposed by Watson et al. allows users of the INTRABEAM (Carl Zeiss Medical AG, Jena, Germany) electronic brachytherapy system to accurately determine the absorbed dose to water, in the absence of a primary dosimetry standard. However, all published values are for PTW 34013 ionization chambers calibrated in a TW30 reference beam, traceable to PTB (Germany). For North American users, it would be advantageous to have data for chambers calibrated in a kV reference beam maintained by the National Institute of Standards and Technology (NIST).

Purpose

In this work, we determine for a PTW 34013 chamber calibrated in three NIST‐traceable reference beams: M30, L40, and L50.

Methods

Using available photon spectra data for M30, L40, and L50 reference beam qualities, Monte Carlo simulations using EGSnrc were performed to calculate the ratio of the absorbed dose to the PTW 34013 chamber air cavity to air‐kerma () for these beams. From this ratio, as a function of depth in water was determined. The effect of the use of a buildup foil was also investigated. An uncertainty analysis considering both the Type A and Type B uncertainties in the calculation of was performed.

Results

The largest difference in was found between L50 and TW30, with a relative decrease of 1.4% (no buildup) to 1.6% (buildup). For M30 and L40, the differences were minimal compared with measurement uncertainties.

Conclusions

We report values for three NIST‐traceable kV reference beams. This study reinforces the feasibility of adapting the Watson et al. methodology using different kV reference beams, facilitating the use of INTRABEAM in North America and ensuring the continuity and accuracy of dosimetry standards in intraoperative radiation therapy.

Subjects

MeSH terms (MEDLINE)

RadiometryMonte Carlo MethodCalibrationX-RaysReference StandardsUncertaintyBrachytherapy

Author keywords

IORTionization chamberkV dosimetry

Authors and institutions

  • Medical Physics Unit, McGill University, Montreal, Quebec, Canada.
  • Department of Radiation Oncology, Miami Cancer Institute, Miami, Florida, USA.
  • Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.

Cite

APA

Watson, P. G. F., Davis, S., & Culberson, W. S. (2024). Technical note: Determination of CQ$C_{Q}$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities. Medical Physics. https://doi.org/10.1002/mp.17345

Vancouver

Watson PGF, Davis S, Culberson WS. Technical note: Determination of CQ$C_{Q}$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities. Medical Physics. 2024. doi:10.1002/mp.17345

BibTeX
@article{Watson2024,
  author = {Watson, Peter G F and Davis, Stephen and Culberson, Wesley S},
  title = {Technical note: Determination of CQ$C_Q$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17345}
}
Start for free →