EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model

Autumn E Walter, Ahtesham U Khan, Larry A DeWerd

Brachytherapy · 2023

DOI: 10.1016/j.brachy.2022.11.010 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Brachytherapy, 2023. Verfasst von Autumn E Walter und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin, Madison, WI; Department of Human Oncology, School of Me…. Hauptthemen laut MEDLINE: Brachytherapy. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The purpose of this work is to provide measured data for the modified TG43 parameters [DeWerd et al.] for the newest, Galden-cooled S7600 Xoft Axxent source model. METHODS: The measurement of radial dose distributions at distances of 1 cm to 4 cm from the source was performed using TLD100 microcubes, EBT3 film, and an Exradin A26 microionization chamber. The overall uncertainty and reproducibility of each dosimeter was evaluated for its use in determining the radial dose function and dose rate conversion coefficient. An acrylic phantom developed in house for previous works was used to measure the polar anisotropy function using TLD100 microcubes at distances of 1 cm, 2 cm, and 5 cm from the source. RESULTS: The Exradin A26 chamber was deemed most suitable for measuring the radial dose function. Values determined had a maximum k = 1 uncertainty of 1.4%. The dose rate conversion coefficient measured with the chamber was found to be 9.33 ± 0.21cGy/hrμGy/min. TLD100 microcube measurements of the polar anisotropy had average uncertainties of 6%, 3%, and 2.5% at 1 cm, 2 cm, and 5 cm, respectively. CONCLUSIONS: The modified TG43 parameters for the bare source were measured with reasonable uncertainty. The values determined will aid with the clinical implementation of the source for breast and endometrial cancer applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyHumansRadiometryReproducibility of ResultsAnisotropyUncertaintyRadiotherapy DosageMonte Carlo Method

Stichwörter der Autor:innen

DosimetryElectronic brachytherapyFilm dosimetryMicroionization chamberModified TG43 parametersTLD100 microcubes

Autor:innen und Institutionen

  • Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin, Madison, WI; Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin, Madison, WI. Electr…
  • Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin, Madison, WI.

Zitieren

APA

Walter, A. E., Khan, A. U., & DeWerd, L. A. (2023). Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model. Brachytherapy. https://doi.org/10.1016/j.brachy.2022.11.010

Vancouver

Walter AE, Khan AU, DeWerd LA. Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model. Brachytherapy. 2023. doi:10.1016/j.brachy.2022.11.010

BibTeX
@article{Walter2023,
  author = {Walter, Autumn E and Khan, Ahtesham U and DeWerd, Larry A},
  title = {Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model},
  journal = {Brachytherapy},
  year = {2023},
  doi = {10.1016/j.brachy.2022.11.010}
}
Kostenlos starten →