Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy

Claas Wessels, Adam Strzelecki, Mathieu Plamondon, Mathias Lehmann, Igor Peterlik, Pascal Paysan, Balazs Nagy, Alexander Heinz, Dieter Seghers, Stephen Thompson, Stefan G Scheib

Medical Physics · 2024

DOI: 10.1002/mp.17325 ↗

At a glance

Evaluationsstudie in Medical Physics, 2024. Written by Claas Wessels and 10 co-authors. Institution: Varian Medical Systems Imaging Laboratory, Daettwil, Switzerland. Major MEDLINE topics: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted. Cited 21 times so far (Crossref); 26 references.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

High‐quality 3D‐anatomy of the day is needed for treatment plan adaptation in radiotherapy. For online x‐ray‐based CBCT workflows, one approach is to create a synthetic CT or to utilize a fan‐beam CT with corresponding registrations. The former potentially introduces uncertainties in the dose calculation if deformable image registration is used. The latter can introduce burden and complexity to the process, the facility, and the patient.

Purpose

Using the CBCT of the day, acquired on the treatment device, for direct dose calculation and plan adaptation can overcome these limitations. This study aims to assess the accuracy of the calculated dose on the CBCT scans acquired on a Halcyon linear accelerator equipped with HyperSight.

Methods

HyperSight's new CBCT reconstruction algorithm includes improvements in scatter correction, HU calibration of the imager, and beam shape adaptation. Furthermore, HyperSight introduced a new x‐ray detector. To show the effect of the implemented improvements, gamma comparisons of 2%/2 mm, 2%/1 mm, and 1%/1 mm were made between the dose distribution in phantoms calculated on the CBCT reconstructions and the simulation CT scans, considering this the standard of care. The resulting gamma passing rates were compared to those obtained with the Halcyon 3.0 reconstruction and hardware without HyperSight's technologies. Various anatomical phantoms for dosimetric evaluations on brain, head and neck, lung, breast, and prostate cases have been used in this study.

Results

The overall results demonstrated that HyperSight outperformed the Halcyon 3.0 version. Based on the gamma analysis, the calculated dose using HyperSight was closer to the CT scan‐based doses than the calculated dose using iCBCT Halcyon 3.0 for most cases. Over all plans and gamma criteria, Halcyon 3.0 achieved an average passing rate of 92.9%, whereas HyperSight achieved 98.1%.

Conclusion

Using HyperSight CBCT images for direct dose calculation, for example, in (online) plan adaptation, seems feasible for the investigated cases.

Subjects

MeSH terms (MEDLINE)

Cone-Beam Computed TomographyPhantoms, ImagingRadiotherapy DosageRadiotherapy Planning, Computer-AssistedHumansRadiation DosageMaleAlgorithmsImage Processing, Computer-Assisted

Author keywords

CBCTadaptive radiation therapydose calculation

Authors and institutions

  • Varian Medical Systems Imaging Laboratory, Daettwil, Switzerland.
  • Varian Medical Systems, Palo Alto, California, USA.

Cite

APA

Wessels, C., Strzelecki, A., Plamondon, M., Lehmann, M., Peterlik, I., Paysan, P., Nagy, B., Heinz, A., Seghers, D., Thompson, S., & Scheib, S. G. (2024). Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy. Medical Physics. https://doi.org/10.1002/mp.17325

Vancouver

Wessels C, Strzelecki A, Plamondon M, Lehmann M, Peterlik I, Paysan P, et al. Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy. Medical Physics. 2024. doi:10.1002/mp.17325

BibTeX
@article{Wessels2024,
  author = {Wessels, Claas and Strzelecki, Adam and Plamondon, Mathieu and Lehmann, Mathias and Peterlik, Igor and Paysan, Pascal and Nagy, Balazs and Heinz, Alexander and Seghers, Dieter and Thompson, Stephen and Scheib, Stefan G},
  title = {Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17325}
}
Start for free →