Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy

Claas Wessels, Adam Strzelecki, Mathieu Plamondon, Mathias Lehmann, Igor Peterlik, Pascal Paysan, Balazs Nagy, Alexander Heinz, Dieter Seghers, Stephen Thompson, Stefan G Scheib

Medical Physics · 2024

DOI: 10.1002/mp.17325 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Medical Physics, 2024. Verfasst von Claas Wessels und 10 weiteren Autor:innen. Beteiligte Einrichtung: Varian Medical Systems Imaging Laboratory, Daettwil, Switzerland. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 26 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

High‐quality 3D‐anatomy of the day is needed for treatment plan adaptation in radiotherapy. For online x‐ray‐based CBCT workflows, one approach is to create a synthetic CT or to utilize a fan‐beam CT with corresponding registrations. The former potentially introduces uncertainties in the dose calculation if deformable image registration is used. The latter can introduce burden and complexity to the process, the facility, and the patient.

Purpose

Using the CBCT of the day, acquired on the treatment device, for direct dose calculation and plan adaptation can overcome these limitations. This study aims to assess the accuracy of the calculated dose on the CBCT scans acquired on a Halcyon linear accelerator equipped with HyperSight.

Methods

HyperSight's new CBCT reconstruction algorithm includes improvements in scatter correction, HU calibration of the imager, and beam shape adaptation. Furthermore, HyperSight introduced a new x‐ray detector. To show the effect of the implemented improvements, gamma comparisons of 2%/2 mm, 2%/1 mm, and 1%/1 mm were made between the dose distribution in phantoms calculated on the CBCT reconstructions and the simulation CT scans, considering this the standard of care. The resulting gamma passing rates were compared to those obtained with the Halcyon 3.0 reconstruction and hardware without HyperSight's technologies. Various anatomical phantoms for dosimetric evaluations on brain, head and neck, lung, breast, and prostate cases have been used in this study.

Results

The overall results demonstrated that HyperSight outperformed the Halcyon 3.0 version. Based on the gamma analysis, the calculated dose using HyperSight was closer to the CT scan‐based doses than the calculated dose using iCBCT Halcyon 3.0 for most cases. Over all plans and gamma criteria, Halcyon 3.0 achieved an average passing rate of 92.9%, whereas HyperSight achieved 98.1%.

Conclusion

Using HyperSight CBCT images for direct dose calculation, for example, in (online) plan adaptation, seems feasible for the investigated cases.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyPhantoms, ImagingRadiotherapy DosageRadiotherapy Planning, Computer-AssistedHumansRadiation DosageMaleAlgorithmsImage Processing, Computer-Assisted

Stichwörter der Autor:innen

CBCTadaptive radiation therapydose calculation

Autor:innen und Institutionen

  • Varian Medical Systems Imaging Laboratory, Daettwil, Switzerland.
  • Varian Medical Systems, Palo Alto, California, USA.

Zitieren

APA

Wessels, C., Strzelecki, A., Plamondon, M., Lehmann, M., Peterlik, I., Paysan, P., Nagy, B., Heinz, A., Seghers, D., Thompson, S., & Scheib, S. G. (2024). Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy. Medical Physics. https://doi.org/10.1002/mp.17325

Vancouver

Wessels C, Strzelecki A, Plamondon M, Lehmann M, Peterlik I, Paysan P, et al. Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy. Medical Physics. 2024. doi:10.1002/mp.17325

BibTeX
@article{Wessels2024,
  author = {Wessels, Claas and Strzelecki, Adam and Plamondon, Mathieu and Lehmann, Mathias and Peterlik, Igor and Paysan, Pascal and Nagy, Balazs and Heinz, Alexander and Seghers, Dieter and Thompson, Stephen and Scheib, Stefan G},
  title = {Technical note: Phantom‐based evaluation of CBCT dose calculation accuracy for use in adaptive radiotherapy},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17325}
}
Kostenlos starten →