A finely segmented semi‐monolithic detector tailored for high‐resolution PET
Medical Physics · 2024
DOI: 10.1002/mp.16928 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Yannick Kuhl und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany. Hauptthemen laut MEDLINE: Positron-Emission Tomography. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 70 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: German Federal Ministry of Education and Research.
Abstract (Originalsprache)
Background
Preclinical research and organ‐dedicated applications use and require high (spatial‐)resolution positron emission tomography (PET) detectors to visualize small structures (early) and understand biological processes at a finer level of detail. Researchers seeking to improve detector and image spatial resolution have explored various detector designs. Current commercial high‐resolution systems often employ finely pixelated or monolithic scintillators, each with its limitations.
Purpose
We present a semi‐monolithic detector, tailored for high‐resolution PET applications with a spatial resolution in the range of 1 mm or better, merging concepts of monolithic and pixelated crystals. The detector features LYSO slabs measuring (24 × 10 × 1) mm 3 , coupled to a 12 × 12 readout channel photosensor with 4 mm pitch. The slabs are grouped in two arrays of 44 slabs each to achieve a higher optical photon density despite the fine segmentation.
Methods
We employ a fan beam collimator for fast calibration to train machine‐learning‐based positioning models for all three dimensions, including slab identification and depth‐of‐interaction (DOI), utilizing gradient tree boosting (GTB). The data for all dimensions was acquired in less than 2 h. Energy calculation was based on a position‐dependent energy calibration. Using an analytical timing calibration, time skews were corrected for coincidence timing resolution (CTR) estimation.
Results
Leveraging machine‐learning‐based calibration in all three dimensions, we achieved high detector spatial resolution: down to 1.18 mm full width at half maximum (FWHM) detector spatial resolution and 0.75 mm mean absolute error (MAE) in the planar‐monolithic direction, and 2.14 mm FWHM and 1.03 mm MAE for DOI at an energy window of (435–585) keV. Correct slab interaction identification in planar‐segmented direction exceeded 80%, alongside an energy resolution of 12.7% and a CTR of 450 ps FWHM.
Conclusions
The introduced finely segmented, high‐resolution slab detector demonstrates appealing performance characteristics suitable for high‐resolution PET applications. The current benchtop‐based detector calibration routine allows these detectors to be used in PET systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany.
- Hyperion Hybrid Imaging Systems GmbH, Aachen, Germany.
- Physics Institute III B, RWTH Aachen University, Aachen, Germany.
ORCID-Profile: Yannick Kuhl, Florian Mueller, Stephan Naunheim, David Schug, Bjoern Weissler, Pierre Gebhardt, Volkmar Schulz
Förderung
- German Federal Ministry of Education and Research
Zitieren
APA
Kuhl, Y., Mueller, F., Naunheim, S., Bovelett, M., Lambertus, J., Schug, D., Weissler, B., Gegenmantel, E., Gebhardt, P., & Schulz, V. (2024). A finely segmented semi‐monolithic detector tailored for high‐resolution PET. Medical Physics. https://doi.org/10.1002/mp.16928
Vancouver
Kuhl Y, Mueller F, Naunheim S, Bovelett M, Lambertus J, Schug D, et al. A finely segmented semi‐monolithic detector tailored for high‐resolution PET. Medical Physics. 2024. doi:10.1002/mp.16928
BibTeX
@article{Kuhl2024,
author = {Kuhl, Yannick and Mueller, Florian and Naunheim, Stephan and Bovelett, Matthias and Lambertus, Janko and Schug, David and Weissler, Bjoern and Gegenmantel, Eike and Gebhardt, Pierre and Schulz, Volkmar},
title = {A finely segmented semi‐monolithic detector tailored for high‐resolution PET},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.16928}
}
EUVIMED