EUVIMEDOpdag EUVIMED
ÜbersichtsarbeitOpen access · CC BY-NC-ND 4.0

A review on 4D cone‐beam CT (4D‐CBCT) in radiation therapy: Technical advances and clinical applications

Yawei Zhang, Zhuoran Jiang, You Zhang, Lei Ren

Medical Physics · 2024-08 · 51 (8) · 5164-5180

DOI: 10.1002/mp.17269 ↗

At a glance

Übersichtsarbeit in Medical Physics, published online on 2024-06-23. Written by Yawei Zhang and 3 co-authors. Institution: Department of Radiation Oncology, University of Florida Health Proton Therapy Institute, Jacksonville, Florida, USA. Major MEDLINE topics: Cone-Beam Computed Tomography, Four-Dimensional Computed Tomography. Cited 40 times so far (Crossref); 150 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: NIBIB NIH HHS, NIH HHS, National Institutes of Health.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Cone‐beam CT (CBCT) is the most commonly used onboard imaging technique for target localization in radiation therapy. Conventional 3D CBCT acquires x‐ray cone‐beam projections at multiple angles around the patient to reconstruct 3D images of the patient in the treatment room. However, despite its wide usage, 3D CBCT is limited in imaging disease sites affected by respiratory motions or other dynamic changes within the body, as it lacks time‐resolved information. To overcome this limitation, 4D‐CBCT was developed to incorporate a time dimension in the imaging to account for the patient's motion during the acquisitions. For example, respiration‐correlated 4D‐CBCT divides the breathing cycles into different phase bins and reconstructs 3D images for each phase bin, ultimately generating a complete set of 4D images. 4D‐CBCT is valuable for localizing tumors in the thoracic and abdominal regions where the localization accuracy is affected by respiratory motions. This is especially important for hypofractionated stereotactic body radiation therapy (SBRT), which delivers much higher fractional doses in fewer fractions than conventional fractionated treatments. Nonetheless, 4D‐CBCT does face certain limitations, including long scanning times, high imaging doses, and compromised image quality due to the necessity of acquiring sufficient x‐ray projections for each respiratory phase. In order to address these challenges, numerous methods have been developed to achieve fast, low‐dose, and high‐quality 4D‐CBCT. This paper aims to review the technical developments surrounding 4D‐CBCT comprehensively. It will explore conventional algorithms and recent deep learning‐based approaches, delving into their capabilities and limitations. Additionally, the paper will discuss the potential clinical applications of 4D‐CBCT and outline a future roadmap, highlighting areas for further research and development. Through this exploration, the readers will better understand 4D‐CBCT's capabilities and potential to enhance radiation therapy.

Subjects

MeSH terms (MEDLINE)

Cone-Beam Computed TomographyFour-Dimensional Computed TomographyHumansRadiotherapy, Image-GuidedRespiration

Author keywords

4D‐CBCTimage reconstructionmotion managementpatient positioning

Authors and institutions

  • Department of Radiation Oncology, University of Florida Health Proton Therapy Institute, Jacksonville, Florida, USA.
  • Department of Radiation Oncology, University of Florida College of Medicine, Gainesville, Florida, USA.
  • Medical Physics Graduate Program, Duke University, Durham, North Carolina, USA.
  • Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA.
  • Department of Radiation Oncology, University of Maryland, Baltimore, Maryland, USA.

Funding

  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Cite

APA

Zhang, Y., Jiang, Z., Zhang, Y., & Ren, L. (2024). A review on 4D cone‐beam CT (4D‐CBCT) in radiation therapy: Technical advances and clinical applications. Medical Physics, 51(8), 5164-5180. https://doi.org/10.1002/mp.17269

Vancouver

Zhang Y, Jiang Z, Zhang Y, Ren L. A review on 4D cone‐beam CT (4D‐CBCT) in radiation therapy: Technical advances and clinical applications. Medical Physics. 2024;51(8):5164-5180. doi:10.1002/mp.17269

BibTeX
@article{Zhang2024,
  author = {Zhang, Yawei and Jiang, Zhuoran and Zhang, You and Ren, Lei},
  title = {A review on 4D cone‐beam CT (4D‐CBCT) in radiation therapy: Technical advances and clinical applications},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {8},
  pages = {5164--5180},
  doi = {10.1002/mp.17269}
}
Start for free →