Ultrafast T 2 and T 2 * mapping using single‐shot spatiotemporally encoded MRI with reduced field of view and spiral out‐in‐out‐in trajectory

Junxian Jin, Yang Zhou, Lin Chen, Zhong Chen

Medical Physics · 2024-10 · 51 (10) · 7308-7319

DOI: 10.1002/mp.17268 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 19.06.2024. Verfasst von Junxian Jin und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, School of Electronic Science and Engineering, Na…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 62 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Shenzhen Science and Technology Program, Fujian Province Science and Technology Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

T 2 and T 2 * mapping are crucial components of quantitative magnetic resonance imaging, offering valuable insights into tissue characteristics and pathology. Single‐shot methods can achieve ultrafast T 2 or T 2 * mapping by acquiring multiple readout echo trains. However, the extended echo trains pose challenges, such as compromised image quality and diminished quantification accuracy.

Purpose

In this study, we develop a single‐shot method for ultrafast T 2 and T 2 * mapping with reduced echo train length.

Methods

The proposed method is based on ultrafast single‐shot spatiotemporally encoded (SPEN) MRI combined with reduced field of view (FOV) and spiral out‐in‐out‐in (OIOI) trajectory. Specifically, a biaxial SPEN excitation scheme was employed to excite the spin signal into the spatiotemporal encoding domain. The OIOI trajectory with high acquisition efficiency was employed to acquire signals within targeted reduced FOV. Through non‐Cartesian super‐resolved (SR) reconstruction, 12 aliasing‐free images with different echo times were obtained within 150 ms. These images were subsequently fitted to generate T 2 or T 2 * mapping simultaneously using a derived model.

Results

Accurate and co‐registered T 2 and T 2 * maps were generated, closely resembling the reference maps. Numerical simulations demonstrated substantial consistency ( R 2 > 0.99) with the ground truth values. A mean difference of 0.6% and 1.7% was observed in T 2 and T 2 *, respectively, in in vivo rat brain experiments compared to the reference. Moreover, the proposed method successfully obtained T 2 and T 2 * mappings of rat kidney in free‐breathing mode, demonstrating its superiority over multishot methods lacking respiratory navigation.

Conclusions

The results suggest that the proposed method can achieve ultrafast and accurate T 2 and T 2 * mapping, potentially facilitating the application of T 2 and T 2 * mapping in scenarios requiring high temporal resolution.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage Processing, Computer-AssistedRatsAnimalsBrainTime FactorsPhantoms, Imaging

Stichwörter der Autor:innen

flexible field of viewquantitative T2 mappingquantitative T2* mappingsingle‐shotspatiotemporally encodedspiral out‐in‐out‐inultrafast

Autor:innen und Institutionen

  • Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, School of Electronic Science and Engineering, National Model Microelectronics College, Xiamen University, Xiamen, Chin…
  • Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.

Förderung

  • National Natural Science Foundation of China
  • Shenzhen Science and Technology Program
  • Fujian Province Science and Technology Project
  • Xiamen University Nanqiang Outstanding Talents Program

Zitieren

APA

Jin, J., Zhou, Y., Chen, L., & Chen, Z. (2024). Ultrafast T 2 and T 2 * mapping using single‐shot spatiotemporally encoded MRI with reduced field of view and spiral out‐in‐out‐in trajectory. Medical Physics, 51(10), 7308-7319. https://doi.org/10.1002/mp.17268

Vancouver

Jin J, Zhou Y, Chen L, Chen Z. Ultrafast T 2 and T 2 * mapping using single‐shot spatiotemporally encoded MRI with reduced field of view and spiral out‐in‐out‐in trajectory. Medical Physics. 2024;51(10):7308-7319. doi:10.1002/mp.17268

BibTeX
@article{Jin2024,
  author = {Jin, Junxian and Zhou, Yang and Chen, Lin and Chen, Zhong},
  title = {Ultrafast T 2 and T 2 * mapping using single‐shot spatiotemporally encoded MRI with reduced field of view and spiral out‐in‐out‐in trajectory},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {10},
  pages = {7308--7319},
  doi = {10.1002/mp.17268}
}