EUVIMEDObjevte EUVIMED
Open access

Fully automatic online geometric calibration for non‐circular cone‐beam CT orbits using fiducials with unknown placement

Yiqun Q Ma iD, Tess Reynolds, Tina Ehtiati, Clifford Weiss, Kelvin Hong, Nicholas Theodore, Grace J Gang, J Webster Stayman

Medical Physics · 2024

DOI: 10.1002/mp.17041 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Yiqun Q Ma and 7 co-authors. Institution: Johns Hopkins University, Baltimore, Maryland, USA. Major MEDLINE topics: Cone-Beam Computed Tomography, Algorithms. Cited 9 times so far (Crossref); 71 references. The full text is freely available in PubMed Central. Funded by: NIBIB NIH HHS, National Institute of Biomedical Imaging and Bioengineering.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

Cone‐beam CT (CBCT) with non‐circular scanning orbits can improve image quality for 3D intraoperative image guidance. However, geometric calibration of such scans can be challenging. Existing methods typically require a prior image, specialized phantoms, presumed repeatable orbits, or long computation time.

Purpose

We propose a novel fully automatic online geometric calibration algorithm that does not require prior knowledge of fiducial configuration. The algorithm is fast, accurate, and can accommodate arbitrary scanning orbits and fiducial configurations.

Methods

The algorithm uses an automatic initialization process to eliminate human intervention in fiducial localization and an iterative refinement process to ensure robustness and accuracy. We provide a detailed explanation and implementation of the proposed algorithm. Physical experiments on a lab test bench and a clinical robotic C‐arm scanner were conducted to evaluate spatial resolution performance and robustness under realistic constraints.

Results

Qualitative and quantitative results from the physical experiments demonstrate high accuracy, efficiency, and robustness of the proposed method. The spatial resolution performance matched that of our existing benchmark method, which used a 3D‐2D registration‐based geometric calibration algorithm.

Conclusions

We have demonstrated an automatic online geometric calibration method that delivers high spatial resolution and robustness performance. This methodology enables arbitrary scan trajectories and should facilitate translation of such acquisition methods in a clinical setting.

Subjects

MeSH terms (MEDLINE)

Cone-Beam Computed TomographyAlgorithmsCalibrationPhantoms, ImagingAutomationHumansFiducial MarkersImaging, Three-Dimensional

Author keywords

CBCTC‐armfiducial configurationgeometric calibrationimage qualityinterventional imagingquality controltask‐driven imaging

Authors and institutions

  • Johns Hopkins University, Baltimore, Maryland, USA.
  • Faculty of Medicine and Health, University of Sydney, Sydney, Australia.
  • Siemens Healthineers, Forchheim, Germany.
  • University of Pennsylvania, Philadelphia, USA.

ORCID profiles: Yiqun Q Ma

Funding

  • NIBIB NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering

Cite

APA

Ma, Y. Q., Reynolds, T., Ehtiati, T., Weiss, C., Hong, K., Theodore, N., Gang, G. J., & Stayman, J. W. (2024). Fully automatic online geometric calibration for non‐circular cone‐beam CT orbits using fiducials with unknown placement. Medical Physics. https://doi.org/10.1002/mp.17041

Vancouver

Ma YQ, Reynolds T, Ehtiati T, Weiss C, Hong K, Theodore N, et al. Fully automatic online geometric calibration for non‐circular cone‐beam CT orbits using fiducials with unknown placement. Medical Physics. 2024. doi:10.1002/mp.17041

BibTeX
@article{Ma2024,
  author = {Ma, Yiqun Q and Reynolds, Tess and Ehtiati, Tina and Weiss, Clifford and Hong, Kelvin and Theodore, Nicholas and Gang, Grace J and Stayman, J Webster},
  title = {Fully automatic online geometric calibration for non‐circular cone‐beam CT orbits using fiducials with unknown placement},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17041}
}
Start for free →