Clinical reference dosimetry for the 0.5 T inline rotating biplanar Linac‐MR
Medical Physics · 2024
DOI: 10.1002/mp.16951 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Eugene Yip und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology, Medical Physics Division, University of Alberta, Edmonton, Alberta, Canada. Hauptthemen laut MEDLINE: Radiometry, Magnetic Fields, Phenylpropionates. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
The world's first clinical 0.5 T inline rotating biplanar Linac‐MR system is commissioned for clinical use. For reference dosimetry, unique features to device, including an SAD = 120 cm, bore clearance of 60 cm × 110 cm, as well as 0.5 T inline magnetic field, provide some challenges to applying a standard dosimetry protocol (i.e., TG‐51).
Purpose
In this work, we propose a simple and practical clinical reference dosimetry protocol for the 0.5T biplanar Linac‐MR and validated its results.
Methods
Our dosimetry protocol for this system is as follows: tissue phantom ratios at 20 and 10 cm are first measured and converted into %dd10x beam quality specifier using equations provided and Kalach and Rogers. The converted %dd10x is used to determine the ion chamber correction factor, using the equations in the TG‐51 addendum for the Exradin A12 farmer chamber used, which is cross‐calibrated with one calibrated at a standards laboratory. For a 0.5 T parallel field, magnetic field effect on chamber response is assumed to have no effect and is not explicitly corrected for. Once the ion chamber correction factor for a non‐standard SAD ( k Q,msr ) is determined, TG‐51 is performed to obtain dose at a depth of 10 cm at SAD = 120 cm. The dosimetry protocol is repeated with the magnetic field ramped down. To validate our dosimetry protocol, Monte Carlo (EGSnrc) simulations are performed to confirm the determined k Q,msr values. MC Simulations and magnetic Field On versus Field Off measurements are performed to confirm that the magnetic field has no effect. To validate our overall dosimetry protocol, external dose audits, based on optical simulated luminescent dosimeters, thermal luminescent dosimeters, and alanine dosimeters are performed on the 0.5 T Linac‐MR system.
Results
Our EGSnrc results confirm our protocol‐determined k Q,msr values, as well as our assumptions about magnetic field effects ( k B = 1) within statistical uncertainty for the A‐12 chamber. Our external dosimetry procedures also validated our overall dosimetry protocol for the 0.5 T biplanar Linac‐MR hybrid. Ramping down the magnetic field has resulted in a dosimetric difference of 0.1%, well within experimental uncertainty.
Conclusion
With the 0.5 T parallel magnetic field having minimal effect on the ion chamber response, a TPR 20,10 approach to determine beam quality provides an accurate method to perform clinical dosimetry for the 0.5 T biplanar Linac‐MR.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Oncology, Medical Physics Division, University of Alberta, Edmonton, Alberta, Canada.
- Department of Medical Physics, Cross Cancer Institute, Edmonton, Alberta, Canada.
- Department of Radiation Oncology, BC Cancer - Victoria, Victoria, British Columbia, Canada.
- Department of Radiaiton Oncology, The Queen's Medical Centre, Honolulu, Hawaii, USA.
- MagnetTx Oncology Solutions, Edmonton, Alberta, Canada.
- Department of Medical Physics, Nova Scotia Health, Halifax, Nova Scotia, Canada.
Zitieren
APA
Yip, E., Tari, S. Y., Reynolds, M. W., Sinn, D., Murray, B. R., Fallone, B. G., & Oliver, P. A. (2024). Clinical reference dosimetry for the 0.5 T inline rotating biplanar Linac‐MR. Medical Physics. https://doi.org/10.1002/mp.16951
Vancouver
Yip E, Tari SY, Reynolds MW, Sinn D, Murray BR, Fallone BG, et al. Clinical reference dosimetry for the 0.5 T inline rotating biplanar Linac‐MR. Medical Physics. 2024. doi:10.1002/mp.16951
BibTeX
@article{Yip2024,
author = {Yip, Eugene and Tari, Shima Y and Reynolds, Michael W and Sinn, David and Murray, Brad R and Fallone, B Gino and Oliver, Patricia Ak},
title = {Clinical reference dosimetry for the 0.5 T inline rotating biplanar Linac‐MR},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.16951}
}
EUVIMED