A finely segmented semi‐monolithic detector tailored for high‐resolution PET

Yannick Kuhl iD, Florian Mueller iD, Stephan Naunheim iD, Matthias Bovelett, Janko Lambertus, David Schug iD, Bjoern Weissler iD, Eike Gegenmantel, Pierre Gebhardt iD, Volkmar Schulz iD

Medical Physics · 2024

DOI: 10.1002/mp.16928 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Yannick Kuhl and 9 co-authors. Institution: Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany. Major MEDLINE topics: Positron-Emission Tomography. Cited 17 times so far (Crossref); 70 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: German Federal Ministry of Education and Research.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

Preclinical research and organ‐dedicated applications use and require high (spatial‐)resolution positron emission tomography (PET) detectors to visualize small structures (early) and understand biological processes at a finer level of detail. Researchers seeking to improve detector and image spatial resolution have explored various detector designs. Current commercial high‐resolution systems often employ finely pixelated or monolithic scintillators, each with its limitations.

Purpose

We present a semi‐monolithic detector, tailored for high‐resolution PET applications with a spatial resolution in the range of 1 mm or better, merging concepts of monolithic and pixelated crystals. The detector features LYSO slabs measuring (24 × 10 × 1) mm 3 , coupled to a 12 × 12 readout channel photosensor with 4 mm pitch. The slabs are grouped in two arrays of 44 slabs each to achieve a higher optical photon density despite the fine segmentation.

Methods

We employ a fan beam collimator for fast calibration to train machine‐learning‐based positioning models for all three dimensions, including slab identification and depth‐of‐interaction (DOI), utilizing gradient tree boosting (GTB). The data for all dimensions was acquired in less than 2 h. Energy calculation was based on a position‐dependent energy calibration. Using an analytical timing calibration, time skews were corrected for coincidence timing resolution (CTR) estimation.

Results

Leveraging machine‐learning‐based calibration in all three dimensions, we achieved high detector spatial resolution: down to 1.18 mm full width at half maximum (FWHM) detector spatial resolution and 0.75 mm mean absolute error (MAE) in the planar‐monolithic direction, and 2.14 mm FWHM and 1.03 mm MAE for DOI at an energy window of (435–585) keV. Correct slab interaction identification in planar‐segmented direction exceeded 80%, alongside an energy resolution of 12.7% and a CTR of 450 ps FWHM.

Conclusions

The introduced finely segmented, high‐resolution slab detector demonstrates appealing performance characteristics suitable for high‐resolution PET applications. The current benchtop‐based detector calibration routine allows these detectors to be used in PET systems.

Subjects

MeSH terms (MEDLINE)

Positron-Emission TomographyEquipment DesignImage Processing, Computer-AssistedCalibration

Author keywords

high‐resolution Positron Emission Tomography (PET)machine learning GTB fan beam calibrationsemi‐monolith slab detector

Authors and institutions

  • Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany.
  • Hyperion Hybrid Imaging Systems GmbH, Aachen, Germany.
  • Physics Institute III B, RWTH Aachen University, Aachen, Germany.

ORCID profiles: Yannick Kuhl, Florian Mueller, Stephan Naunheim, David Schug, Bjoern Weissler, Pierre Gebhardt, Volkmar Schulz

Funding

  • German Federal Ministry of Education and Research

Cite

APA

Kuhl, Y., Mueller, F., Naunheim, S., Bovelett, M., Lambertus, J., Schug, D., Weissler, B., Gegenmantel, E., Gebhardt, P., & Schulz, V. (2024). A finely segmented semi‐monolithic detector tailored for high‐resolution PET. Medical Physics. https://doi.org/10.1002/mp.16928

Vancouver

Kuhl Y, Mueller F, Naunheim S, Bovelett M, Lambertus J, Schug D, et al. A finely segmented semi‐monolithic detector tailored for high‐resolution PET. Medical Physics. 2024. doi:10.1002/mp.16928

BibTeX
@article{Kuhl2024,
  author = {Kuhl, Yannick and Mueller, Florian and Naunheim, Stephan and Bovelett, Matthias and Lambertus, Janko and Schug, David and Weissler, Bjoern and Gegenmantel, Eike and Gebhardt, Pierre and Schulz, Volkmar},
  title = {A finely segmented semi‐monolithic detector tailored for high‐resolution PET},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16928}
}
Start for free →