Multiscale Monte Carlo simulations for dosimetry in x‐ray breast imaging: Part II ‐ Microscopic scales
Medical Physics · 2024
DOI: 10.1002/mp.16912 ↗At a glance
Journal article in Medical Physics, 2024. Written by Rodrigo T Massera and 2 co-authors. Institution: Instituto de Física Gleb Wataghin, Universidade Estadual de Campinas (UNICAMP), Campinas, São Paulo, Brazil. Major MEDLINE topics: Radiometry, Mammography. Cited 3 times so far (Crossref); 29 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: Fundação de Amparo à Pesquisa do Estado de São Paulo, Conselho Nacional de Desenvolvimento Científico e Tecnológico, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior.
Abstract (original language)
Background
Although the benefits of breast screening and early diagnosis are known for reducing breast cancer mortality rates, the effects and risks of low radiation doses to the cells in the breast are still ongoing topics of study.
Purpose
To study specific energy distributions () in cytoplasm and nuclei of cells corresponding to glandular tissue for different x‐ray breast imaging modalities.
Methods
A cubic lattice (500 μm length side) containing 4064 spherical cells was irradiated with photons loaded from phase space files with varying glandular voxel doses (). Specific energy distributions were scored for nucleus and cytoplasm compartments using the PENELOPE (v. 2018) + penEasy (v. 2020) Monte Carlo (MC) code. The phase space files, generated in part I of this work, were obtained from MC simulations in a voxelized anthropomorphic phantom corresponding to glandular voxels for different breast imaging modalities, including digital mammography (DM), digital breast tomosynthesis (DBT), contrast enhanced digital mammography (CEDM) and breast CT (BCT).
Results
In general, the average specific energy in nuclei is higher than the respective glandular dose scored in the same region, by up to 10%. The specific energy distributions for nucleus and cytoplasm are directly related to the magnitude of the glandular dose in the voxel (), with little dependence on the spatial location. For similar values, for nuclei is different between DM/DBT and CEDM/BCT, indicating that distinct x‐ray spectra play significant roles in . In addition, this behavior is also present when the specific energy distribution () is considered taking into account the GDD in the breast.
Conclusions
Microdosimetry studies are complementary to the traditional macroscopic breast dosimetry based on the mean glandular dose (MGD). For the same MGD, the specific energy distribution in glandular tissue varies between breast imaging modalities, indicating that this effect could be considered for studying the risks of exposing the breast to ionizing radiation.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Instituto de Física Gleb Wataghin, Universidade Estadual de Campinas (UNICAMP), Campinas, São Paulo, Brazil.
- Carleton Laboratory for Radiotherapy Physics, Department of Physics, Carleton University, Ottawa, Ontario, Canada.
Funding
- Fundação de Amparo à Pesquisa do Estado de São Paulo
- Conselho Nacional de Desenvolvimento Científico e Tecnológico
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
- Natural Sciences and Engineering Research Council of Canada
- Emerging leaders of the Americas program of the Government of Canada
- Canada Research Charis
- Fundo de Apoio ao Ensino, a Pesquisa e Extensão da UNICAMP
Cite
APA
Massera, R. T., Tomal, A., & Thomson, R. M. (2024). Multiscale Monte Carlo simulations for dosimetry in x‐ray breast imaging: Part II ‐ Microscopic scales. Medical Physics. https://doi.org/10.1002/mp.16912
Vancouver
Massera RT, Tomal A, Thomson RM. Multiscale Monte Carlo simulations for dosimetry in x‐ray breast imaging: Part II ‐ Microscopic scales. Medical Physics. 2024. doi:10.1002/mp.16912
BibTeX
@article{Massera2024,
author = {Massera, Rodrigo T and Tomal, Alessandra and Thomson, Rowan M},
title = {Multiscale Monte Carlo simulations for dosimetry in x‐ray breast imaging: Part II ‐ Microscopic scales},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.16912}
}