Systematic evaluation of human soft tissue attenuation correction in whole‐body PET/MR: Implications from PET/CT for optimization of MR‐based AC in patients with normal lung tissue

Maike E Lindemann, Marcel Gratz, Hong Grafe, Kai Jannusch, Lale Umutlu, Harald H Quick

Medical Physics · 2024

DOI: 10.1002/mp.16863 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Maike E Lindemann and 5 co-authors. Institution: High-Field and Hybrid MR Imaging, University Hospital Essen, University Duisburg-Essen, Essen, Germany. Major MEDLINE topics: Positron Emission Tomography Computed Tomography, Multimodal Imaging. Cited 4 times so far (Crossref); 49 references. The full text is freely available (open access, CC BY-NC-ND 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Attenuation correction (AC) is an important methodical step in positron emission tomography/magnetic resonance imaging (PET/MRI) to correct for attenuated and scattered PET photons.

Purpose

The overall quality of magnetic resonance (MR)‐based AC in whole‐body PET/MRI was evaluated in direct comparison to computed tomography (CT)‐based AC serving as reference. The quantitative impact of isolated tissue classes in the MR‐AC was systematically investigated to identify potential optimization needs and strategies.

Methods

Data of n = 60 whole‐body PET/CT patients with normal lung tissue and without metal implants/prostheses were used to generate six different AC‐models based on the CT data for each patient, simulating variations of MR‐AC. The original continuous CT‐AC (CT‐org) is referred to as reference. A pseudo MR‐AC (CT‐mrac), generated from CT data, with four tissue classes and a bone atlas represents the MR‐AC. Relative difference in linear attenuation coefficients (LAC) and standardized uptake values were calculated. From the results two improvements regarding soft tissue AC and lung AC were proposed and evaluated.

Results

The overall performance of MR‐AC is in good agreement compared to CT‐AC. Lungs, heart, and bone tissue were identified as the regions with most deviation to the CT‐AC (myocardium −15%, bone tissue −14%, and lungs ±20%). Using single‐valued LACs for AC in the lung only provides limited accuracy. For improved soft tissue AC, splitting the combined soft tissue class into muscles and organs each with adapted LAC could reduce the deviations to the CT‐AC to < ±1%. For improved lung AC, applying a gradient LAC in the lungs could remarkably reduce over‐ or undercorrections in PET signal compared to CT‐AC (±5%).

Conclusions

The AC is important to ensure best PET image quality and accurate PET quantification for diagnostics and radiotherapy planning. The optimized segment‐based AC proposed in this study, which was evaluated on PET/CT data, inherently reduces quantification bias in normal lung tissue and soft tissue compared to the CT‐AC reference.

Subjects

MeSH terms (MEDLINE)

Positron Emission Tomography Computed TomographyMultimodal ImagingHumansImage Processing, Computer-AssistedMagnetic Resonance ImagingMagnetic Resonance SpectroscopyTomography, X-Ray ComputedPositron-Emission TomographyLung

Author keywords

AC in PET/CTattenuation correction (AC)improved soft tissue AClung AC in PET/MRquantitative PET/MRwhole-body PET/MR

Authors and institutions

  • High-Field and Hybrid MR Imaging, University Hospital Essen, University Duisburg-Essen, Essen, Germany.
  • Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, Germany.
  • Department of Nuclear Medicine, University Hospital Essen, University Duisburg-Essen, Essen, Germany.
  • Department of Diagnostic and Interventional Radiology, University Hospital Duesseldorf, University Duesseldorf, Duesseldorf, Germany.
  • Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.

Cite

APA

Lindemann, M. E., Gratz, M., Grafe, H., Jannusch, K., Umutlu, L., & Quick, H. H. (2024). Systematic evaluation of human soft tissue attenuation correction in whole‐body PET/MR: Implications from PET/CT for optimization of MR‐based AC in patients with normal lung tissue. Medical Physics. https://doi.org/10.1002/mp.16863

Vancouver

Lindemann ME, Gratz M, Grafe H, Jannusch K, Umutlu L, Quick HH. Systematic evaluation of human soft tissue attenuation correction in whole‐body PET/MR: Implications from PET/CT for optimization of MR‐based AC in patients with normal lung tissue. Medical Physics. 2024. doi:10.1002/mp.16863

BibTeX
@article{Lindemann2024,
  author = {Lindemann, Maike E and Gratz, Marcel and Grafe, Hong and Jannusch, Kai and Umutlu, Lale and Quick, Harald H},
  title = {Systematic evaluation of human soft tissue attenuation correction in whole‐body PET/MR: Implications from PET/CT for optimization of MR‐based AC in patients with normal lung tissue},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16863}
}
Start for free →