EUVIMEDOpdag EUVIMED
Crossref

Proof‐of‐concept for a thin conical X‐ray target optimized for intensity and directionality for use in a carbon nanotube‐based compact X‐ray tube

Ben Insley, Dirk Bartkoski, Peter Balter, Surendra Prajapati, Ramesh Tailor, Mohammad Salehpour, David Jaffray

Medical Physics · 2024

DOI: 10.1002/mp.16835 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Ben Insley and 6 co-authors. Institution: Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Major MEDLINE topics: Nanotubes, Carbon. Cited 3 times so far (Crossref); 40 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Carbon nanotube‐based cold cathode technology has revolutionized the miniaturization of X‐ray tubes. However, current applications of these devices required optimization for large, uniform fields with low intensity.

Purpose

This work investigated the feasibility and radiological characteristics of a novel conical X‐ray target optimized for high intensity and high directionality to be used in a compact X‐ray tube.

Methods

The proposed device uses an ultrathin, conical tungsten‐diamond target that exhibits significant heat loading while maintaining a small focal spot size and promoting forward‐directedness of the X‐ray field through preferential attenuation of oblique‐angled photons. The electrostatic and thermal properties of the theoretical tube were calculated and analyzed using COMSOL Multiphysics software. The production, transport, and calculation of radiological properties associated with the resultant X‐ray field were performed using the Geant4 toolkit via its wrapper, TOPAS.

Results

Heat transfer analysis of this X‐ray tube demonstrated the feasibility of a 200‐kV electron beam bombarding the proposed target at a maximum current of 100 mA using a 1‐ms symmetric duty cycle. The cathode of the X‐ray tube was designed to be segmented into nine switchable electrical segments for modulation of the focal spot size from 0.4‐ to 10.8‐mm. After importing the COMSOL‐derived electron beam into TOPAS for X‐ray production simulations, radiological analysis of the resultant field demonstrated high levels of intrinsic beam collimation while maintaining high intensity. A maximum dose rate of 17,887 cGy/min was calculated for 1‐mm depth in water at 7‐cm distance.

Conclusions

The proposed X‐ray tube design can create highly directional X‐ray fields with superior fluence compared to that of current commercial X‐ray tubes of comparable size.

Subjects

MeSH terms (MEDLINE)

Nanotubes, CarbonX-RaysRadiographyFluoroscopySoftwareMonte Carlo Method

Author keywords

Monte Carlocarbon nanotubesminiature X-ray tubeproof-of-concept

Substances

Nanotubes, Carbon

Authors and institutions

  • Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Empyrean Medical Systems, Inc., 950 Peninsula Corp Cir, Boca Raton, USA.
  • Division of Office of the Sr. VP & Chief Technology and Digital Officer, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.

Cite

APA

Insley, B., Bartkoski, D., Balter, P., Prajapati, S., Tailor, R., Salehpour, M., & Jaffray, D. (2024). Proof‐of‐concept for a thin conical X‐ray target optimized for intensity and directionality for use in a carbon nanotube‐based compact X‐ray tube. Medical Physics. https://doi.org/10.1002/mp.16835

Vancouver

Insley B, Bartkoski D, Balter P, Prajapati S, Tailor R, Salehpour M, et al. Proof‐of‐concept for a thin conical X‐ray target optimized for intensity and directionality for use in a carbon nanotube‐based compact X‐ray tube. Medical Physics. 2024. doi:10.1002/mp.16835

BibTeX
@article{Insley2024,
  author = {Insley, Ben and Bartkoski, Dirk and Balter, Peter and Prajapati, Surendra and Tailor, Ramesh and Salehpour, Mohammad and Jaffray, David},
  title = {Proof‐of‐concept for a thin conical X‐ray target optimized for intensity and directionality for use in a carbon nanotube‐based compact X‐ray tube},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16835}
}
Start for free →