Characterization of a 0.8 mm 3 Medscint plastic scintillator detector system for small field dosimetry
Physics in Medicine & Biology · 2023
DOI: 10.1088/1361-6560/aceacf ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2023. Verfasst von Elena Timakova und 2 weiteren Autor:innen. Beteiligte Einrichtung: University of Victoria, British Columbia, Canada. Hauptthemen laut MEDLINE: Radiometry, Particle Accelerators. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective . Plastic scintillator detectors (PSDs) have demonstrated ability to meet requirements of small field dosimetry. Medscint developed a 1 mm long, 1 mm diameter cylindrical PSD with effective volume of 0.8 mm 3 . Clinically relevant, small field dosimetric properties of this detector, combined with a novel scintillation dosimetry system—HYPERSCINT RP-200, and HYPERDOSE analysis software were evaluated in this study. Approach . This novel scintillator-based dosimetry system was characterized with 6 MV-WFF and 10 MV-FFF x-ray beams delivered by Varian TrueBeam TM linear accelerator. The detector was characterized for leakage, short-term repeatability, dose response linearity, angular response, dose rate response, and field size dependence for radiation field sizes of 0.25 × 0.25 to 10 × 10 cm 2 . Measured detector specific output ratios were compared with microDiamond output factors to determine small field output correction factors, k Q clin , Q msr f clin , f msr . Main results . The dosimetry system showed excellent short-term repeatability with standard deviation of only 0.04 ± 0.01%. It demonstrated good dose linearity with variations less than 1.0% for 14.4 cGy and above. The dosimetry system was found to be independent of dose rate and angle of irradiation, with deviations for both below 0.5%. Leakage was found to be comparable to background readings. For 6 MV-WFF energy beams, detector specific output ratios for field sizes down to 1 × 1 cm 2 agreed with output factors measured with PTW TN60019 microDiamond, thus, k Q clin , Q msr f clin , f msr equates to unity for these field sizes. For 10 MV-FFF energy beams, detector specific output ratios for field sizes down to 2 × 2 cm 2 agreed with PTW TN60019 microDiamond output factors, thus, k Q clin , Q msr f clin , f msr equates to unity for these field sizes. k Q clin , Q msr f clin , f msr for field sizes down to 0.5 × 0.5 cm 2 were determined to be within 6% of unity for both 6 MV-WFF and 10 MV-FFF energy beams. Significance . The HYPERSCINT RP-200 dosimetry system coupled with a 0.8 mm 3 PSD showed excellent dosimetric properties and was found to be clinically relevant for relative dosimetry down to field sizes of 0.5 × 0.5 cm 2 and potentially smaller.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Victoria, British Columbia, Canada.
- BC Cancer Agency, Vancouver Island Centre, British Columbia, Canada.
Zitieren
APA
Timakova, E., Bazalova-Carter, M., & Zavgorodni, S. (2023). Characterization of a 0.8 mm 3 Medscint plastic scintillator detector system for small field dosimetry. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/aceacf
Vancouver
Timakova E, Bazalova-Carter M, Zavgorodni S. Characterization of a 0.8 mm 3 Medscint plastic scintillator detector system for small field dosimetry. Physics in Medicine & Biology. 2023. doi:10.1088/1361-6560/aceacf
BibTeX
@article{Timakova2023,
author = {Timakova, Elena and Bazalova-Carter, Magdalena and Zavgorodni, Sergei},
title = {Characterization of a 0.8 mm 3 Medscint plastic scintillator detector system for small field dosimetry},
journal = {Physics in Medicine & Biology},
year = {2023},
doi = {10.1088/1361-6560/aceacf}
}
EUVIMED