EUVIMEDOpdag EUVIMED
Multizentrische StudieOpen access · CC BY 4.0

IAEA/WHO postal dosimetry audit methodology for electron beams using radio photoluminescent dosimeters

Alexis Dimitriadis, Pavel Kazantsev, Krzysztof Chelminski, Egor Titovich, Ekaterina Naida, Talent Magnus, Ahmed Meghzifene, Godfrey Azangwe, Mauro Carrara, Jamema Swamidas

Medical Physics · 2023

DOI: 10.1002/mp.16776 ↗

At a glance

Multizentrische Studie in Medical Physics, 2023. Written by Alexis Dimitriadis and 9 co-authors. Institution: Dosimetry and Medical Radiation Physics Section, Division of Human Health, Department of Nuclear Sciences and Applications, International Atomic Ener…. Major MEDLINE topics: Radiation Dosimeters, Electrons. Cited 15 times so far (Crossref); 34 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Independent dosimetry audits are an important intervention in radiotherapy for quality assurance. Electron beams, used for superficial radiotherapy treatments, must also be tested in dosimetry audits as part of a good quality assurance program to help prevent clinical errors.

Purpose

To establish a new service for IAEA/WHO postal dosimetry audits in electron beams using RPL dosimeters.

Methods

A novel postal audit methodology employing a PMMA holder system for RPLDs was developed. The associated correction factors including holder dependence, energy dependence, dose response non‐linearity, and fading were obtained and tested in a multi‐center ( n = 12) pilot study. A measurement uncertainty budget was estimated and employed in analyzing the irradiated dosimeters.

Results

Holder and energy correction factors ranged from 1.004 to 1.010 and 1.019 to 1.059 respectively across the energy range. The non‐linearity and fading correction models used for photon beams were tested in electron beams and did not significantly increase measurement uncertainty. The mean dose ratio ± SD of the multi‐center study was 1.001 ± 0.011. The overall uncertainty budget was estimated as ± 1.42% ( k = 1).

Conclusions

A methodology for IAEA/WHO postal dosimetry audits in electron beams was developed and validated in a multi‐center study and is now made available to radiotherapy centers as a routine service.

Subjects

MeSH terms (MEDLINE)

Radiation DosimetersElectronsRadiotherapy DosagePilot ProjectsThermoluminescent DosimetryRadiometryWorld Health Organization

Author keywords

auditdosimetryelectron

Authors and institutions

  • Dosimetry and Medical Radiation Physics Section, Division of Human Health, Department of Nuclear Sciences and Applications, International Atomic Energy Agency, Vienna, Austria.

Cite

APA

Dimitriadis, A., Kazantsev, P., Chelminski, K., Titovich, E., Naida, E., Magnus, T., Meghzifene, A., Azangwe, G., Carrara, M., & Swamidas, J. (2023). IAEA/WHO postal dosimetry audit methodology for electron beams using radio photoluminescent dosimeters. Medical Physics. https://doi.org/10.1002/mp.16776

Vancouver

Dimitriadis A, Kazantsev P, Chelminski K, Titovich E, Naida E, Magnus T, et al. IAEA/WHO postal dosimetry audit methodology for electron beams using radio photoluminescent dosimeters. Medical Physics. 2023. doi:10.1002/mp.16776

BibTeX
@article{Dimitriadis2023,
  author = {Dimitriadis, Alexis and Kazantsev, Pavel and Chelminski, Krzysztof and Titovich, Egor and Naida, Ekaterina and Magnus, Talent and Meghzifene, Ahmed and Azangwe, Godfrey and Carrara, Mauro and Swamidas, Jamema},
  title = {IAEA/WHO postal dosimetry audit methodology for electron beams using radio photoluminescent dosimeters},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16776}
}
Start for free →