Multizentrische Studie in Medical Physics, 2023. Written by Alexis Dimitriadis and 9 co-authors. Institution: Dosimetry and Medical Radiation Physics Section, Division of Human Health, Department of Nuclear Sciences and Applications, International Atomic Ener…. Major MEDLINE topics: Radiation Dosimeters, Electrons. Cited 15 times so far (Crossref); 34 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
Independent dosimetry audits are an important intervention in radiotherapy for quality assurance. Electron beams, used for superficial radiotherapy treatments, must also be tested in dosimetry audits as part of a good quality assurance program to help prevent clinical errors.
Purpose
To establish a new service for IAEA/WHO postal dosimetry audits in electron beams using RPL dosimeters.
Methods
A novel postal audit methodology employing a PMMA holder system for RPLDs was developed. The associated correction factors including holder dependence, energy dependence, dose response non‐linearity, and fading were obtained and tested in a multi‐center ( n = 12) pilot study. A measurement uncertainty budget was estimated and employed in analyzing the irradiated dosimeters.
Results
Holder and energy correction factors ranged from 1.004 to 1.010 and 1.019 to 1.059 respectively across the energy range. The non‐linearity and fading correction models used for photon beams were tested in electron beams and did not significantly increase measurement uncertainty. The mean dose ratio ± SD of the multi‐center study was 1.001 ± 0.011. The overall uncertainty budget was estimated as ± 1.42% ( k = 1).
Conclusions
A methodology for IAEA/WHO postal dosimetry audits in electron beams was developed and validated in a multi‐center study and is now made available to radiotherapy centers as a routine service.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Dosimetry and Medical Radiation Physics Section, Division of Human Health, Department of Nuclear Sciences and Applications, International Atomic Energy Agency, Vienna, Austria.
Cite
APA
Dimitriadis, A., Kazantsev, P., Chelminski, K., Titovich, E., Naida, E., Magnus, T., Meghzifene, A., Azangwe, G., Carrara, M., & Swamidas, J. (2023). IAEA/WHO postal dosimetry audit methodology for electron beams using radio photoluminescent dosimeters. Medical Physics. https://doi.org/10.1002/mp.16776
Vancouver
Dimitriadis A, Kazantsev P, Chelminski K, Titovich E, Naida E, Magnus T, et al. IAEA/WHO postal dosimetry audit methodology for electron beams using radio photoluminescent dosimeters. Medical Physics. 2023. doi:10.1002/mp.16776
BibTeX
@article{Dimitriadis2023,
author = {Dimitriadis, Alexis and Kazantsev, Pavel and Chelminski, Krzysztof and Titovich, Egor and Naida, Ekaterina and Magnus, Talent and Meghzifene, Ahmed and Azangwe, Godfrey and Carrara, Mauro and Swamidas, Jamema},
title = {IAEA/WHO postal dosimetry audit methodology for electron beams using radio photoluminescent dosimeters},
journal = {Medical Physics},
year = {2023},
doi = {10.1002/mp.16776}
}
EUVIMED