EUVIMEDEUVIMED entdecken
Crossref

Contour subregion error detection methodology using deep learning auto‐segmentation

Jingwei Duan, Mark E Bernard, Yi Rong, James R Castle, Xue Feng, Jeremiah D Johnson, Quan Chen

Medical Physics · 2023

DOI: 10.1002/mp.16768 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Jingwei Duan und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky, USA. Hauptthemen laut MEDLINE: Deep Learning, Head and Neck Neoplasms. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: NCI NIH HHS, Varian Medical Systems.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Inaccurate manual organ delineation is one of the high‐risk failure modes in radiation treatment. Numerous automated contour quality assurance (QA) systems have been developed to assess contour acceptability; however, manual inspection of flagged cases is a time‐consuming and challenging process, and can lead to users overlooking the exact error location.

Purpose

Our aim is to develop and validate a contour QA system that can effectively detect and visualize subregional contour errors, both qualitatively and quantitatively. Methods/Materials A novel contour subregion error detection (CSED) system was developed using subregional surface distance discrepancies between manual and deep learning auto‐segmentation (DLAS) contours. A validation study was conducted using a head and neck public dataset containing 339 cases and evaluated according to knowledge‐based pass criteria derived from a clinical training dataset of 60 cases. A blind qualitative evaluation was conducted, comparing the results from the CSED system with manual labels. Subsequently, the CSED‐flagged cases were re‐examined by a radiation oncologist.

Results

The CSED system could visualize the diverse types of subregional contour errors qualitatively and quantitatively. In the validation dataset, the CSED system resulted in true positive rates (TPR) of 0.814, 0.800, and 0.771; false positive rates (FPR) of 0.310, 0.267, and 0.298; and accuracies of 0.735, 0.759, and 0.730, for brainstem and left and right parotid contours, respectively. The CSED‐assisted manual review caught 13 brainstem, 19 left parotid, and 21 right parotid contour errors missed by conventional human review. The TPR/FPR/accuracy of the CSED‐assisted manual review improved to 0.836/0.253/0.784, 0.831/0.171/0.830, and 0.808/0.193/0.807 for each structure, respectively. Further, the time savings achieved through CSED‐assisted review improved by 75%, with the time for review taking 24.81 ± 12.84, 26.75 ± 10.41, and 28.71 ± 13.72 s for each structure, respectively.

Conclusions

The CSED system enables qualitative and quantitative detection, localization, and visualization of manual segmentation subregional errors utilizing DLAS contours as references. The use of this system has been shown to help reduce the risk of high‐risk failure modes resulting from inaccurate organ segmentation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningHead and Neck NeoplasmsHumansRadiotherapy Planning, Computer-AssistedTomography, X-Ray ComputedNeckOrgans at RiskImage Processing, Computer-Assisted

Stichwörter der Autor:innen

OAR delineationdeep learning auto segmentationquality assurance

Autor:innen und Institutionen

  • Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky, USA.
  • Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.
  • Carina Medical LLC, Lexington, Kentucky, USA.
  • Department of Radiation Oncology, City of Hope Comprehensive Cancer Center, Duarte, California, USA.

Förderung

  • NCI NIH HHS
  • Varian Medical Systems

Zitieren

APA

Duan, J., Bernard, M. E., Rong, Y., Castle, J. R., Feng, X., Johnson, J. D., & Chen, Q. (2023). Contour subregion error detection methodology using deep learning auto‐segmentation. Medical Physics. https://doi.org/10.1002/mp.16768

Vancouver

Duan J, Bernard ME, Rong Y, Castle JR, Feng X, Johnson JD, et al. Contour subregion error detection methodology using deep learning auto‐segmentation. Medical Physics. 2023. doi:10.1002/mp.16768

BibTeX
@article{Duan2023,
  author = {Duan, Jingwei and Bernard, Mark E and Rong, Yi and Castle, James R and Feng, Xue and Johnson, Jeremiah D and Chen, Quan},
  title = {Contour subregion error detection methodology using deep learning auto‐segmentation},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16768}
}
Kostenlos starten →