Technical note: MR‐compatible pedal ergometer with electromechanical pedal resistance and exercise triggering enhanced by visual feedback via video display
Medical Physics · 2023
DOI: 10.1002/mp.16682 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Petr Sedivy und 8 weiteren Autor:innen. Beteiligte Einrichtung: MR Unit, Department of Diagnostic and Interventional Radiology, Institute for Clinical and Experimental Medicine, Prague, Czech Republic. Hauptthemen laut MEDLINE: Feedback, Sensory, Magnetic Resonance Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Health, Czech Republic-DRO "Institute for Clinical and Experimental Medicine - IKEM, IN 00023001", National Institute for Research of Metabolic and Cardiovascular Diseases. Funded by the European Union - Next Generation EU, Programme EXCELES.
Abstract (Originalsprache)
Background
During and after exercise, dynamic 31 P MR parameters are typically measured using an MR‐compatible ergometer. Self‐built equipment for local condition can be constructed where possible.
Purpose
To develop a pedal resistance ergometer with rocker arm based on a system that combines electric weight displacement, visual self‐monitoring, and exercise triggering. The repeatability and reproducibility were tested.
Methods
The hardware and software for the ergometer were constructed from commercial components in a home laboratory. Twelve volunteers participated in the testing of the ergometer.
Results
A fully automated ergometer system was developed, allowing the pedal resistance to be adjusted during the examination. The system includes a self‐monitoring and triggering mechanism that enables both the operator and subject to monitor pedal frequency and force. The operator can modify the pedal resistance as desired during the exercise. This self‐monitoring solution is simple and cost‐effective, requiring only a commercial potentiometer, an Arduino converter, and a conventional video projector with a personal computer (PC). Additionally, all system components are located outside the magnetic resonance (MR) room, avoiding interference with the MR system. Results of several test of the reproducibility/repeatability of power at three pedal resistance values (15%, 24%, 25% maximal voluntary force) were expressed both as a coefficient of variation ranging from 6% to 3.1% and as an intraclass correlation of coefficient ranging from 0.96 to 0.99. Similar values were also found for other dynamic parameters of 31 P MR spectroscopy. These findings are similar to published data obtained on different types of ergometers.
Conclusions
Based on more than 1 year of usage, the ergometer proved successful in handling stationary and variable loads, and can be easily operated by a single user.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- MR Unit, Department of Diagnostic and Interventional Radiology, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
- Information Technology Division, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
- Department of Medical Technology and Investments, Institute for Clinical and Experimental Medicine, Prague, Czech Republic.
- ICAR, Prague, Czech Republic.
Förderung
- Ministry of Health, Czech Republic-DRO "Institute for Clinical and Experimental Medicine - IKEM, IN 00023001"
- National Institute for Research of Metabolic and Cardiovascular Diseases. Funded by the European Union - Next Generation EU, Programme EXCELES
Zitieren
APA
Sedivy, P., Dezortova, M., Rydlo, J., Moravec, P., Krizek, I., Setinova, B., Pajuelo, D., Burian, M., & Hajek, M. (2023). Technical note: MR‐compatible pedal ergometer with electromechanical pedal resistance and exercise triggering enhanced by visual feedback via video display. Medical Physics. https://doi.org/10.1002/mp.16682
Vancouver
Sedivy P, Dezortova M, Rydlo J, Moravec P, Krizek I, Setinova B, et al. Technical note: MR‐compatible pedal ergometer with electromechanical pedal resistance and exercise triggering enhanced by visual feedback via video display. Medical Physics. 2023. doi:10.1002/mp.16682
BibTeX
@article{Sedivy2023,
author = {Sedivy, Petr and Dezortova, Monika and Rydlo, Jan and Moravec, Petr and Krizek, Ivan and Setinova, Bara and Pajuelo, Dita and Burian, Martin and Hajek, Milan},
title = {Technical note: MR‐compatible pedal ergometer with electromechanical pedal resistance and exercise triggering enhanced by visual feedback via video display},
journal = {Medical Physics},
year = {2023},
doi = {10.1002/mp.16682}
}
EUVIMED