Simultaneous Assessment of Skeletal Muscle Energetics and Blood Flow During Dynamic Exercise by Interleaved 31 P ‐ MRS / 1 H ‐ MRI
Magnetic Resonance in Medicine · 2026-07 · 96 (1) · 13-26
DOI: 10.1002/mrm.70337 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 17.03.2026. Verfasst von T Jake Samuel und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Magnetic Resonance Research, Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medic…. Hauptthemen laut MEDLINE: Muscle, Skeletal, Exercise, Magnetic Resonance Imaging. Die Arbeit nennt 65 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIA NIH HHS, NIH HHS, NHLBI NIH HHS.
Abstract (Originalsprache)
Purpose
To simultaneously measure skeletal muscle energetics and blood flow (BF) before, during, and after dynamic plantar flexion exercise (PFE).
Methods
Non‐localized pulse‐acquire phosphorus‐31 magnetic resonance spectroscopy ( 31 P MRS) and phase contrast flow magnetic resonance imaging ( 1 H MRI) using golden‐angle rotated spiral readouts were acquired in an interleaved fashion. Data were collected at rest, during PFE to volitional fatigue, and post‐exercise recovery in 10 healthy adults. Interleaved popliteal artery BF was validated against conventional cine phase‐contrast measures at rest and post‐exercise recovery, while calf muscle energetics were validated by comparing to conventional 31 P MRS‐only measures at rest. To minimize motion artifacts during dynamic PFE, flow data were resolved in 3 dimensions (cardiac phase, pedal position, and exercise stage).
Results
Resting phosphocreatine (PCr) and inorganic phosphate concentrations were similar between both techniques (both p > 0.05). During PFE, PCr declined to 46% ± 10% of rest. PCr recovery time was 43 ± 15 s. Resting BF was 132 ± 43 mL/min and increased up to four‐fold during PFE depending on exercise stage and pedal position. Notably, there was excellent agreement between the interleaved and conventional BF measures ( r 2 = 0.95, slope = 0.977, p < 0.0001) with minimal bias (12.62 mL/min) over a wide range of physiologically relevant BF values (from ∼60 to 600 mL/min).
Conclusion
The feasibility of simultaneous assessment of skeletal muscle energetics and BF during dynamic exercise by interleaved 31 P MRS/ 1 H flow MRI was demonstrated along with good agreement between interleaved and conventional techniques during rest and recovery. One can now simultaneously quantify critical BF‐metabolism relationships and use this to probe disease‐related adaptations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Division of Magnetic Resonance Research, Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
- North America Clinical Science, MR R&D, Philips, Cambridge, Massachusetts, USA.
- Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota, USA.
- Division of Cardiology, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
ORCID-Profile: T Jake Samuel, Sandeep K Ganji, Joseph R Goldenberg, Sabra C Lewsey, Allison G Hays, Robert G Weiss, Michael Schär
Förderung
- NIA NIH HHS
- NIH HHS
- NHLBI NIH HHS
- Clarence Doodeman Endowment in Cardiology at Johns Hopkins University
- Johns Hopkins Older Americans Independence Center Junior Faculty Investigator Award from the Johns Hopkins University Claude D. Pepper Older Americans Independ…
- National Institutes of Health
Zitieren
APA
Samuel, T. J., Ganji, S. K., Goldenberg, J. R., Lewsey, S. C., Hays, A. G., Weiss, R. G., & Schär, M. (2026). Simultaneous Assessment of Skeletal Muscle Energetics and Blood Flow During Dynamic Exercise by Interleaved 31 P ‐ MRS / 1 H ‐ MRI. Magnetic Resonance in Medicine, 96(1), 13-26. https://doi.org/10.1002/mrm.70337
Vancouver
Samuel TJ, Ganji SK, Goldenberg JR, Lewsey SC, Hays AG, Weiss RG, et al. Simultaneous Assessment of Skeletal Muscle Energetics and Blood Flow During Dynamic Exercise by Interleaved 31 P ‐ MRS / 1 H ‐ MRI. Magnetic Resonance in Medicine. 2026;96(1):13-26. doi:10.1002/mrm.70337
BibTeX
@article{Samuel2026,
author = {Samuel, T Jake and Ganji, Sandeep K and Goldenberg, Joseph R and Lewsey, Sabra C and Hays, Allison G and Weiss, Robert G and Schär, Michael},
title = {Simultaneous Assessment of Skeletal Muscle Energetics and Blood Flow During Dynamic Exercise by Interleaved 31 P ‐ MRS / 1 H ‐ MRI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {1},
pages = {13--26},
doi = {10.1002/mrm.70337}
}