Creating patient‐specific digital phantoms with a longitudinal atlas for evaluating deformable CT‐CBCT registration in adaptive lung radiotherapy
Medical Physics · 2024
DOI: 10.1002/mp.16606 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Sebastian Meyer und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA. Hauptthemen laut MEDLINE: Spiral Cone-Beam Computed Tomography. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Background
Quality assurance of deformable image registration (DIR) is challenging because the ground truth is often unavailable. In addition, current approaches that rely on artificial transformations do not adequately resemble clinical scenarios encountered in adaptive radiotherapy.
Purpose
We developed an atlas‐based method to create a variety of patient‐specific serial digital phantoms with CBCT‐like image quality to assess the DIR performance for longitudinal CBCT imaging data in adaptive lung radiotherapy.
Methods
A library of deformations was created by extracting the longitudinal changes observed between a planning CT and weekly CBCT from an atlas of lung radiotherapy patients. The planning CT of an inquiry patient was first deformed by mapping the deformation pattern from a matched atlas patient, and subsequently appended with CBCT artifacts to imitate a weekly CBCT. Finally, a group of digital phantoms around an inquiry patient was produced to simulate a series of possible evolutions of tumor and adjacent normal structures. We validated the generated deformation vector fields (DVFs) to ensure numerically and physiologically realistic transformations. The proposed framework was applied to evaluate the performance of the DIR algorithm implemented in the commercial Eclipse treatment planning system in a retrospective study of eight inquiry patients.
Results
The generated DVFs were inverse consistent within less than 3 mm and did not exhibit unrealistic folding. The deformation patterns adequately mimicked the observed longitudinal anatomical changes of the matched atlas patients. Worse Eclipse DVF accuracy was observed in regions of low image contrast or artifacts. The structure volumes exhibiting a DVF error magnitude of equal or more than 2 mm ranged from 24.5% (spinal cord) to 69.2% (heart) and the maximum DVF error exceeded 5 mm for all structures except the spinal cord. Contour‐based evaluations showed a high degree of alignment with dice similarity coefficients above 0.8 in all cases, which underestimated the overall DVF accuracy within the structures.
Conclusions
It is feasible to create and augment digital phantoms based on a particular patient of interest using multiple series of deformation patterns from matched patients in an atlas. This can provide a semi‐automated procedure to complement the quality assurance of CT‐CBCT DIR and facilitate the clinical implementation of image‐guided and adaptive radiotherapy that involve longitudinal CBCT imaging studies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
Förderung
- NCI NIH HHS
Zitieren
APA
Meyer, S., Alam, S., Kuo, L. C., Hu, Y. C., Liu, Y., Lu, W., Yorke, E., Li, A., Cerviño, L., & Zhang, P. (2024). Creating patient‐specific digital phantoms with a longitudinal atlas for evaluating deformable CT‐CBCT registration in adaptive lung radiotherapy. Medical Physics. https://doi.org/10.1002/mp.16606
Vancouver
Meyer S, Alam S, Kuo LC, Hu YC, Liu Y, Lu W, et al. Creating patient‐specific digital phantoms with a longitudinal atlas for evaluating deformable CT‐CBCT registration in adaptive lung radiotherapy. Medical Physics. 2024. doi:10.1002/mp.16606