Technical note: Accuracy and precision of T2 and T2* with a gradient‐echo spin‐echo (GESE) sequence for cardiac imaging

Qi Huang, Jason Mendes, Ganesh Adluru, Edward DiBella

Medical Physics · 2023

DOI: 10.1002/mp.16569 ↗

At a glance

Journal article in Medical Physics, 2023. Written by Qi Huang and 3 co-authors. Institution: Utah Center for Advanced Imaging Research (UCAIR), Department of Radiology and Imaging Sciences, University of Utah, Salt Lake City, Utah, USA. Major MEDLINE topics: Echo-Planar Imaging, Heart. 29 references. Funded by: NHLBI NIH HHS.

TiivistelmäSubjectsBibliografiset tiedotCiteRelated studies

Abstract (original language)

Background

The use of a gradient echo spin echo (GESE) method to obtain rapid T2 and T2* estimation in the heart has been proposed. The effect of acquisition parameter settings on T2 and T2* bias and precision have not been investigated in depth.

Purpose

To understand factors impacting the quantification of T2 and T2* values with a gradient echo spin echo (GESE) method using echo planar imaging (EPI) readouts in a reduced field of view acquisition.

Methods

The GESE method is implemented with a reduced field‐of‐view using an outer volume suppression (OVS) technique to minimize the time for multi‐echo EPI readouts. The number of EPI readouts (images) for the GESE is optimized using Cramer‐Rao Lower Bound (CRLB) and Monte Carlo simulations with a nonlinear least‐square (NLLS) estimator. The SNR requirements were studied using the latter simulation method for a selected range of T2 and T2* values and T2/T2* ratios. Two healthy control subjects were imaged with the proposed GESE sequence and evaluated with the NLLS estimation method. In addition, the proposed OVS method was compared with a saturation bands OVS method in one subject. Clinical T2 and T2* mappings were used as the reference.

Results

The optimal number of EPI readouts is five and the performance is slightly better when the refocusing pulse is placed between the 2 nd and 3 rd readouts. The SNR requirement for achieving a target bias < 1 ms and standard deviation (SD) < 5 ms is more demanding when T2/T2* ratio increases. The minimum SNR requirement in the GESE acquisition should vary from 6 to 20 depending on specific myocardial T2 and T2* values at 3T. The T2 and T2* estimates using the proposed OVS method and the saturation bands OVS method are both similar to the reference.

Conclusion

The GESE sequence with five EPI readouts is a feasible and efficient technique that can estimate T2 and T2* values in the septal myocardium within a heartbeat when the SNR requirement can be satisfied.

Subjects

MeSH terms (MEDLINE)

Echo-Planar ImagingHeartHumansMyocardiumCardiac Imaging TechniquesComputer SimulationMagnetic Resonance Imaging

Author keywords

Cramer-Rao lower boundSNRT2T2*gradient echo and spin echo

Authors and institutions

  • Utah Center for Advanced Imaging Research (UCAIR), Department of Radiology and Imaging Sciences, University of Utah, Salt Lake City, Utah, USA.
  • Department of Biomedical Engineering, University of Utah, Salt Lake City, Utah, USA.

Funding

  • NHLBI NIH HHS

Cite

APA

Huang, Q., Mendes, J., Adluru, G., & DiBella, E. (2023). Technical note: Accuracy and precision of T2 and T2* with a gradient‐echo spin‐echo (GESE) sequence for cardiac imaging. Medical Physics. https://doi.org/10.1002/mp.16569

Vancouver

Huang Q, Mendes J, Adluru G, DiBella E. Technical note: Accuracy and precision of T2 and T2* with a gradient‐echo spin‐echo (GESE) sequence for cardiac imaging. Medical Physics. 2023. doi:10.1002/mp.16569

BibTeX
@article{Huang2023,
  author = {Huang, Qi and Mendes, Jason and Adluru, Ganesh and DiBella, Edward},
  title = {Technical note: Accuracy and precision of T2 and T2* with a gradient‐echo spin‐echo (GESE) sequence for cardiac imaging},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16569}
}
Start for free →