Monte Carlo‐derived 99m Tc uptake quantification with commercial planar MBI: Tumor and breast activity concentrations

Benjamin P Lopez, S Cheenu Kappadath iD

Medical Physics · 2023

DOI: 10.1002/mp.16213 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Benjamin P Lopez und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Neoplasms. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Current molecular breast imaging (MBI) images are limited to qualitative evaluation, not absolute measurement, of 99m Tc uptake in benign and malignant breast tissues.

Purpose

This work assesses the accuracy of previously‐published and newly‐proposed tumor and normal breast tissue 99m Tc uptake MBI measurements using simulations of a commercial dual‐headed planar MBI system under typical clinical and acquisition protocols.

Methods

Quantification techniques were tested in over 4000 simulated acquisitions of spherical and ellipsoid tumors with clinically relevant uptake conditions using a validated Monte Carlo application of the GE Discovery NM750b system. The evaluated techniques consisted of four tumor total activity methodologies (two single‐detector‐based and two geometric‐mean‐based), two tumor MBI volume methodologies (diameter‐based and ROI‐based), and two normal tissue activity concentration methodologies (single‐detector‐based and geometric‐mean‐based). The most accurate of these techniques were then used to estimate tumor activity concentrations and tumor to normal tissue relative activity concentrations (RC).

Results

Single‐detector techniques for tumor total activity quantification achieved mean (standard deviation) relative errors of 0.2% (4.3%) and 1.6% (4.4%) when using the near and far detector images, respectively and were more accurate and precise than the measured 8.1% (5.8%) errors of a previously published geometric‐mean technique. Using these activity estimates and the true tumor volumes resulted in tumor activity concentration and RC errors within 10% of simulated values. The precision of tumor activity concentration and RC when using only MBI measurements were largely driven by the errors in estimating tumor MBI volume using planar images (± 30% inter‐quartile range).

Conclusions

Planar MBI images were shown to accurately and reliably be used to estimate tumor total activities and normal tissue activity concentrations in this simulation study. However, volumetric tumor uptake measurements (i.e., absolute and relative concentrations) are limited by inaccuracies in MBI volume estimation using two‐dimensional images, highlighting the need for either tomographic MBI acquisitions or anatomical volume estimates for accurate three‐dimensional tumor uptake estimates.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

NeoplasmsHumansRadionuclide ImagingBreastMammographyComputer SimulationPhantoms, Imaging

Stichwörter der Autor:innen

absolute quantificationmolecular breast imaging

Autor:innen und Institutionen

  • Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, USA.

ORCID-Profile: S Cheenu Kappadath

Förderung

  • NCI NIH HHS

Zitieren

APA

Lopez, B. P., & Kappadath, S. C. (2023). Monte Carlo‐derived 99m Tc uptake quantification with commercial planar MBI: Tumor and breast activity concentrations. Medical Physics. https://doi.org/10.1002/mp.16213

Vancouver

Lopez BP, Kappadath SC. Monte Carlo‐derived 99m Tc uptake quantification with commercial planar MBI: Tumor and breast activity concentrations. Medical Physics. 2023. doi:10.1002/mp.16213

BibTeX
@article{Lopez2023,
  author = {Lopez, Benjamin P and Kappadath, S Cheenu},
  title = {Monte Carlo‐derived 99m Tc uptake quantification with commercial planar MBI: Tumor and breast activity concentrations},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16213}
}
Kostenlos starten →