IEEE Transactions on Medical Imaging · 2024

DOI: 10.1109/tmi.2023.3307004 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Medical Imaging, 2024. Verfasst von Raymond J Acciavatti und 6 weiteren Autor:innen. Hauptthemen laut MEDLINE: Image Processing, Computer-Assisted, Calcinosis. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIBIB NIH HHS, Department of Defense Breast Cancer Research Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Our lab at the University of Pennsylvania (UPenn) is investigating novel designs for digital breast tomosynthesis. We built a next-generation tomosynthesis system with a non-isocentric geometry (superior-to-inferior detector motion). This paper examines four metrics of image quality affected by this design. First, aliasing was analyzed in reconstructions prepared with smaller pixelation than the detector. Aliasing was assessed with a theoretical model of r -factor, a metric calculating amplitudes of alias signal relative to input signal in the Fourier transform of the reconstruction of a sinusoidal object. Aliasing was also assessed experimentally with a bar pattern (illustrating spatial variations in aliasing) and 360°-star pattern (illustrating directional anisotropies in aliasing). Second, the point spread function (PSF) was modeled in the direction perpendicular to the detector to assess out-of-plane blurring. Third, power spectra were analyzed in an anthropomorphic phantom developed by UPenn and manufactured by Computerized Imaging Reference Systems (CIRS), Inc. (Norfolk, VA). Finally, calcifications were analyzed in the CIRS Model 020 BR3D Breast Imaging Phantom in terms of signal-to-noise ratio (SNR); i.e., mean calcification signal relative to background-tissue noise. Image quality was generally superior in the non-isocentric geometry: Aliasing artifacts were suppressed in both theoretical and experimental reconstructions prepared with smaller pixelation than the detector. PSF width was also reduced at most positions. Anatomic noise was reduced. Finally, SNR in calcification detection was improved. (A potential trade-off of smaller-pixel reconstructions was reduced SNR; however, SNR was still improved by the detector-motion acquisition.) In conclusion, the non-isocentric geometry improved image quality in several ways.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Image Processing, Computer-AssistedCalcinosisHumansBreastMammographyComputer SimulationModels, TheoreticalPhantoms, ImagingAlgorithms

Förderung

  • NCI NIH HHS
  • NIBIB NIH HHS
  • Department of Defense Breast Cancer Research Program
  • Burroughs Wellcome Fund
  • Susan G. Komen®
  • National Institute of Health
  • 2020 American Association of Physicists in Medicine (AAPM) Research Seed Funding Grant
  • 2022 Young Investigator Grant from Breast Cancer Alliance Inc

Zitieren

APA

Acciavatti, R. J., Choi, C. J., Vent, T. L., Barufaldi, B., Cohen, E. A., Wileyto, E. P., & Maidment, A. D. A. (2024). Non-Isocentric Geometry for Next-Generation Tomosynthesis With Super-Resolution. IEEE Transactions on Medical Imaging. https://doi.org/10.1109/tmi.2023.3307004

Vancouver

Acciavatti RJ, Choi CJ, Vent TL, Barufaldi B, Cohen EA, Wileyto EP, et al. Non-Isocentric Geometry for Next-Generation Tomosynthesis With Super-Resolution. IEEE Transactions on Medical Imaging. 2024. doi:10.1109/tmi.2023.3307004

BibTeX
@article{Acciavatti2024,
  author = {Acciavatti, Raymond J and Choi, Chloe J and Vent, Trevor L and Barufaldi, Bruno and Cohen, Eric A and Wileyto, E Paul and Maidment, Andrew D A},
  title = {Non-Isocentric Geometry for Next-Generation Tomosynthesis With Super-Resolution},
  journal = {IEEE Transactions on Medical Imaging},
  year = {2024},
  doi = {10.1109/tmi.2023.3307004}
}
Kostenlos starten →